System Overview
OPMI LUMERA® 700
Version 7.3
Page 78
G-30-1673-en
7
Lamp housing
The light source(s) in the lamp housing supplies/supply light to the SCI il-
lumination system and the additional illuminator (e.g. VISULUX fiber slit il-
luminator). The device can be equipped with one of the following light
source combinations:
Superlux Eye (standard)
The system comes with an integrated lamp housing containing the Su-
perlux Eye light source. The lamp housing features fiber optic illumination,
a motorized filter wheel and two xenon lamps. The second xenon lamp is
used as a backup lamp which can be manually swung into the illumination
beam path if the first xenon lamp fails.
The xenon lamps generate light whose spectrum resembles that of natural
daylight. Regardless of the brightness setting, the color temperature of
the light always remains the same. This permits normal daylight film
without any additional conversion filters to be used for photographic doc-
umentation. The following filters can be swung into the beam path via the
5.7" control panel, the foot control panel or the handgrips:
– Retina protection filter (blue barrier filter). The filter is integrated as a
standard feature. It protects the patient's eye against unnecessary ra-
diation and permits the radiation exposure time to be increased by fac-
tor 5.
– HaMode filter. The filter is integrated as a standard feature and gen-
erates a light spectrum similar to that of a halogen light source.
– Fluorescence filter, 485 nm. This filter makes fluorescent areas visible
and is available as an option.
If the SuperLux Eye light source is equipped with a 485 nm fluorescence
filter, the front panel is marked by the number 304977-9116-500.
Main light source
Additional light
source
Easy
Xenon (Superlux Eye)
-
Halogen (optional)
-
LED (optional)
-
Dual (optional)
Xenon (Superlux Eye)
Halogen
Halogen
Xenon (Superlux Eye)
Halogen
LED
LED
LED
LED
Halogen
LED
Xenon (Superlux Eye)
Содержание opmi lumera 700
Страница 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Страница 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Страница 25: ...Version 7 3 G 30 1673 en Page 25 OPMI LUMERA 700 Safety Measures Fig 1 Safety Equipment 1 4 2 6 7 5 3 8...
Страница 31: ...Version 7 3 G 30 1673 en Page 31 OPMI LUMERA 700 Safety Measures Fig 4 Labels on the microscope 7 5 6 8 1 2 3 4...
Страница 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Страница 49: ...Version 7 3 G 30 1673 en Page 49 OPMI LUMERA 700 System Overview Fig 17 System overview 6 7 2 1 3 5 4...
Страница 57: ...Version 7 3 G 30 1673 en Page 57 OPMI LUMERA 700 System Overview Fig 21 Components of the surgical microscope 5...
Страница 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Страница 71: ...Version 7 3 G 30 1673 en Page 71 OPMI LUMERA 700 System Overview Fig 29 Controls of XY coupling 1...
Страница 89: ...Version 7 3 G 30 1673 en Page 89 OPMI LUMERA 700 System Overview Fig 39 Display and control panel 1 2 3 1 2...
Страница 103: ...Version 7 3 G 30 1673 en Page 103 OPMI LUMERA 700 System Overview Fig 46 Ports on the 22 TFT 1 2 3 4 5 6...
Страница 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Страница 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Страница 111: ...Version 7 3 G 30 1673 en Page 111 OPMI LUMERA 700 Preparations for use Fig 47 Transport position of the 22 TFT 2 3 1 4...
Страница 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...
Страница 115: ...Version 7 3 G 30 1673 en Page 115 OPMI LUMERA 700 Preparations for use Fig 49 Connecting the system 2 5 1 3 4...
Страница 137: ...Version 7 3 G 30 1673 en Page 137 OPMI LUMERA 700 Preparations for use Fig 64 Connecting the foot control panel 3 1 2...