System Overview
OPMI LUMERA® 700
Version 7.3
Page 52
G-30-1673-en
Components of the surgical microscope
1
Tube for main and assistant's microscope
The following tubes can be attached:
– Invertertube - manually adjustable (optional)
The tube can be tilted by 110°. It is particularly suitable for use in com-
bination with a fundus viewing system, as it erects an inverted image.
– Invertertube E - electrically adjustable (standard)
The tube can be tilted by 110°. It is particularly suitable for use in com-
bination with a fundus viewing system, as it erects an inverted image.
– 180° tiltable tube (optional)
This tube can be turned by 180° and permits surgeons to sit upright
even if the microscope is tilted.
– Inclined tube with a 45° viewing angle (optional)
The tube cannot be tilted and is therefore mainly suitable for tasks per-
formed with the microscope in a vertical position.
2
Integrated SCI (Stereo Coaxial Illumination)
The SCI illumination has been specially designed for use in ophthalmology
and is integrated as a standard feature in the surgical microscope. It is
used to illuminate the surgical field and can be controlled via the 5.7" con-
trol panel (see page 224), the foot control panel or handgrips.
There are three fixed SCI settings that can be activated directly or between
which continuously variable adjustment is possible.
Red reflex illumination
Light emitted by the surgical microscope is limited to a diameter of ap-
prox. 20 mm and generates a red reflex which optimally visualizes the
structure of the patient's eye.
Red reflex / surrounding field illumination
The light emitted by the surgical microscope generates a red reflex and si-
multaneously illuminates the surrounding area of the patient's eye. The
ratio is freely configurable (seepage 192)
Surrounding field illumination
The red reflex is deactivated. The entire field of view is illuminated by sur-
rounding field illumination.
Содержание opmi lumera 700
Страница 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Страница 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Страница 25: ...Version 7 3 G 30 1673 en Page 25 OPMI LUMERA 700 Safety Measures Fig 1 Safety Equipment 1 4 2 6 7 5 3 8...
Страница 31: ...Version 7 3 G 30 1673 en Page 31 OPMI LUMERA 700 Safety Measures Fig 4 Labels on the microscope 7 5 6 8 1 2 3 4...
Страница 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Страница 49: ...Version 7 3 G 30 1673 en Page 49 OPMI LUMERA 700 System Overview Fig 17 System overview 6 7 2 1 3 5 4...
Страница 57: ...Version 7 3 G 30 1673 en Page 57 OPMI LUMERA 700 System Overview Fig 21 Components of the surgical microscope 5...
Страница 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Страница 71: ...Version 7 3 G 30 1673 en Page 71 OPMI LUMERA 700 System Overview Fig 29 Controls of XY coupling 1...
Страница 89: ...Version 7 3 G 30 1673 en Page 89 OPMI LUMERA 700 System Overview Fig 39 Display and control panel 1 2 3 1 2...
Страница 103: ...Version 7 3 G 30 1673 en Page 103 OPMI LUMERA 700 System Overview Fig 46 Ports on the 22 TFT 1 2 3 4 5 6...
Страница 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Страница 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Страница 111: ...Version 7 3 G 30 1673 en Page 111 OPMI LUMERA 700 Preparations for use Fig 47 Transport position of the 22 TFT 2 3 1 4...
Страница 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...
Страница 115: ...Version 7 3 G 30 1673 en Page 115 OPMI LUMERA 700 Preparations for use Fig 49 Connecting the system 2 5 1 3 4...
Страница 137: ...Version 7 3 G 30 1673 en Page 137 OPMI LUMERA 700 Preparations for use Fig 64 Connecting the foot control panel 3 1 2...