88
Maintenance / Further information
G-30-1416-en
OPMI® 1FR XY Surgical Microscope on S7 Floor Stand
Issue 4.0
Printed on 14. 11. 2007
Foot control panel
Insufficient surgical field illu-
mination.
Brightness level set too low. Adjust brightness on the
suspension system's dis-
play panel or using the foot
control panel.
Halogen lamp not properly
plugged into lamp mount.
Properly push halogen lamp
into lamp mount.
Defective light guide (illumi-
nation not uniform).
Contact service dept.
Light guide probably needs
to be changed.
-
Motorized focusing system
of microscope inoperative.
Failure of suspension
system electronics.
Using the suspension arm,
set correct focal plane.
Contact service dept.
-
Motion of suspension arm is
too stiff.
Friction adjustment knob
tightened too firmly.
Slightly loosen friction ad-
justment knob.
Stand wobbles.
Floor not level. Stand base
not appropriately posi-
tioned.
Slightly change the stand
base position.
–
Problem
Possible cause
Remedy
See
Problem
Possible cause
Remedy
See
No function of foot control
panel.
Connectors on suspension
system not plugged in.
Plug in connectors.
Содержание OPMI 1FR XY
Страница 6: ...G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 22: ...22 Safety G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 38: ...38 Description G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 106: ...106 Technical data G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 112: ...112 Index G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 113: ...G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007 113...