Lamp housing
The stand is equipped with an illumination system for light guides. The
lamp housing contains a backup lamp which is automatically swung into
the illumination beam path when the first lamp fails.
1
Lamp module
2
Ventilation grid
Do not cover the ventilation grid! Make sure that drapes do not cover
the grid, as this can lead to overheating of the lamp modules and to
lamp failure.
3
Flap
The flap is the mechanical indicator of the operating status of the hal-
ogen lamps:
–
When the flap is closed, the main lamp is operative.
–
When the flap is open, the main lamp has failed. The backup lamp
is on.
4
Manual selection of the backup lamp
If the automatic selector system fails, press this button to switch on the
backup lamp.
5
Opening the lamp module
When you press this button, the lamp module is slightly ejected. Pull
out the lamp module all the way for lamp change.
6
Filter knob
The filter knob has four positions:
0
no filter
1
GG 475 filter: When operating on the eye, always use this filter
to protect the patient's eye against unnecessary (blue) radiation
(retinal injury).
2
KK 40 filter: to increase color temperature
3
no filter
Содержание OPMI 1FR XY
Страница 6: ...G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 22: ...22 Safety G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 38: ...38 Description G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 106: ...106 Technical data G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 112: ...112 Index G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 113: ...G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007 113...