Electromagnetic compatibility
000000-1520-928_GA_EN_021013
57
Guidance and manufacturer’s declaration - electromagnetic immunity
The i.Profiler
® plus
is intended for use in the electromagnetic environment specified below.
The customer or the user of the i.Profiler
® plus
should assure that the device is used in such an environment.
Immunity checks
IEC 60601-1-2 test
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Fast transient
electrical
disturbances/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality should be
that of typical commercial or
hospital environments.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be
that of typical commercial or
hospital environments.
Voltage dips, short
interruptions to the
power supply and
voltage variation on
power supply input
lines
IEC 61000-4-11
< 5 % U
T
(> 95 % dip in U
T
)
for 1/2 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 s
< 5 % U
T
(> 95 % dip in U
T
)
for 1/2 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 s
Mains power quality should be
that of typical commercial or
hospital environments. If the user
of the
i.Profiler
® plus
requires
continued operation during power
mains interruptions, it is
recommended that the
i.Profiler
® plus
be powered from an
uninterruptible power supply or a
battery.
Note: U
T
is the a.c. mains voltage prior to application of the test level.