Installation
000000-1520-928_GA_EN_021013
19
Installation
WARNING - GENERAL HAZARDS
The device may not be stored or operated in ambient conditions other than
those prescribed (see
on page 54).
The device should be set up so that the power cable can be disconnected
from the power supply quickly and easily without any tools.
WARNING - RISK OF ELECTRIC SHOCK
Additional extension cables or portable multiple sockets may not be used.
The electrical installation must conform to IEC 60364-7-710 or the
applicable national regulations. This includes the existence of a ground fault
circuit interrupter (GFCI).
To avert the risk of an electric shock this device may only be connected to a
power supply with a protective earth conductor.
Ensure that the power supply plug is suitable and certified for the local
connection. If the supplied power cable must be replaced, at least the
following specifications must be adhered to:
•
Protective earth conductor resistance maximum 0.2 Ohm
•
Local certification of the power cable for connection to medical devices
•
Device plug C19 conforming to IEC 60320
•
Cross-section at least 0.75 mm²/AWG 18
Hospital Grade design for
specific countries (e.g. USA, Canada)
(For cables > 2.5 m the cross-
section must be increased to 1.5 mm²)
WARNING - FIRE HAZARD
The device is not suitable for operation in explosion risk areas (e.g.
combustible mixture of anesthetic, cleaning, or disinfecting agents with air,
oxygen, or nitrous oxide).
The electrical installation must conform to IEC 60364-7-710.
The data concerning power consumption on the type plate must be taken
into account in the selection of overcurrent protection.