Maintenance and care
000000-1520-928_GA_EN_021013
52
Maintenance
Replacing the print paper roll
WARNING - RISK OF ELECTRIC SHOCK
Disconnect the power cable from the power supply before replacing the
printing paper.
•
Open the paper compartment at the right hand side of the device
(
8
, Fig. 9, page 22).
•
Remove the empty paper roll and insert a new roll according to the
drawing on the inside of the compartment.
•
Pull out the end of the print paper roll so that it remains visible when the
cover has been closed.
•
Close the printing paper compartment.
Care and cleaning
WARNING - RISK OF ELECTRIC SHOCK
Prevent moisture from penetrating the instrument or keyboard. Disconnect
the power cable from the power supply before cleaning or disinfecting the
device.
CAUTION - RISK DUE TO CROSS-CONTAMINATION
Parts with which the patient has come into contact during the examination
(chin rest, forehead rest) should be cleaned with a disinfectant approved for
the purpose. These parts are resistant to wiping off with cleaning agents
categorized as "low" (e.g. suds, quaternary ammonium compounds) and
"intermediate" (e.g. alcohol, Javel water, iodine); classification pursuant to:
Disinfectants and activity spectrum according to the Center for Disease
Control and Prevention, Atlanta, USA).