START-UP
Colibri.2
Construction of the illumination system
Carl Zeiss
07/2014
423052-7244-001
21
3.4.4
External white light source
At the front of the white light source (illumination
system HXP 120 V) there are
−
the
Power
switch (Fig. 3-10/
1
),
−
the
Brightness
turning knob (Fig. 3-10/
2
) with
LED bar graph,
−
the
Shutter
button (Fig. 3-10/
3
) with the
Shutter open
indicator (Fig. 3-10/
4
),
−
the
Lamp Status
indicator (Fig. 3-10/
5
),
−
and the
Confirm Lamp Change
button
(Fig. 3-10/
6
).
After the lamp has fired, the
Lamp Status
indicator shows the state of the internal operation-
hour counter and therefore the age of the lamp in
different colors:
Lamp Status indicator
Internal operation-hour counter
Comment
Green
< 2000 hours
Yellow
> 2000 hours
Prepare lamp change
Red
> 2500 hours
Change lamp
Flashing red
Lamp defective
At the rear there are the power connection socket
(Fig. 3-11/
2
) and the fiber optics outlet
(Fig. 3-11/
3
).
Besides that, the trigger cable port
Trigger in
(Fig. 3-11/
4
) is situated here. The trigger cable
must be used to connect the HXP 120 V to the
Colibri.2 control unit, if the shutter should be
controlled via the Colibri.2 unit. If the shutter
should be controlled by the microscope, the
microscope has to be connected to the
CAN
port
(Fig. 3-11/
1
) of the HXP 120 V.
If the HXP 120 V is used without the
Colibri.2 lamp module, an illumination
adapter is required to couple the liquid
light guide into the reflected light path
of the microscope (423302-0000-000
illumination adapter for light sources
with light guide).
Fig. 3-10
Illumination system HXP 120 V,
front
Fig. 3-11
Illumination system HXP 120 V, rear