OPERATION
Colibri.2
Functions of the white light source HXP 120 V
Carl Zeiss
07/2014
423052-7244-001
23
The display changes according to whether the
lamp module or the external white light source
HXP 120 V is used.
With the lamp module active, the wavelength,
switching status and the power of the installed LED
module is shown (Fig. 4-3/
1
).
Is the white light source active, a display of
Ext
Light Source
is shown in line 1. The shutter
position of the white light source is symbolized
graphically in the lines 2 to 4 (shutter opened: Fig.
4-3/
2
; shutter closed: Fig. 4-3/
3
).
In the event that the upper part of the housing of
the lamp module is opened while the device is
operating, the LED modules will be switched off
automatically. A corresponding message will be
displayed on the control panel.
4.3
Functions of the white light source
HXP 120 V
For operating the white light source please pay
attention to the manufacturer's separate operating
manual as well.
Prior to switching on the white light source please
make sure that the liquid light guide has been
properly positioned and fixed to the Colibri.2 lamp
module.
•
Switch on (or off) the external white light
source via the
Power
switch (Fig. 4-4/
1
).
•
Open or close the shutter by pushing the
Shutter
button (Fig. 4-4/
3
). The
Shutter open
indicator (Fig. 4-4/
4
) shows the operating state
of the shutter.
•
The brightness of the white light source can be
adjusted manually by actuating the turning the
knob
Brightness
(Fig. 4-4/
2
).
If the trigger cable has been connected to the
Colibri.2 control unit, the shutter can also be
opened/closed via the control panel.
The control panel, however, cannot be used to
regulate the intensity of the white light source.
To change over from white light source to LED
modules and vice versa, the
Ext.
button located at
the control panel has to be pressed down.
Fig. 4-3
Various indications on the control
panel display when using the
Colibri.2 illumination system and
the white light source HXP 120 V,
respectively
Fig. 4-4
Illumination system HXP 120 V,
front