MODIFYING THE LAMP MODULE
Carl Zeiss
Removing the lamp module from the microscope and opening it
Colibri.2
24
423052-7244-001
07/2014
5
MODIFYING THE LAMP MODULE
Unplug the power plug of the microscope, the control unit and the connected white light
source from the power supply before mounting/dismounting the lamp module to or from the
microscope and before carrying out any modification! Disconnect the power plug of the control
unit as well before changing the fuses!
Make sure that only fuses for the rated power are used. Use of makeshift fuses and short-
circuiting of the fuse holders are not permitted.
The lamp module is provided with an interlock switch that will disconnect the illumination
system immediately as soon as the upper part of the housing is removed.
5.1
Removing the lamp module from
the microscope and opening it
•
Hold the lamp module firmly, loosen the
clamping screw at the microscope coupling or
at the illumination adapter (Fig. 5-1/
2
) using the
SW 3 ball-headed screwdriver and remove the
lamp module.
•
Loosen both fastening screws (Fig. 5-1/
1
) at the
lamp module.
•
Place the lamp module safely on a suitable base.
•
Pull off the upper part 1 of the housing (Fig.
5-2/
1
) upwards (the upper part 2 (Fig. 5-2/
2
) is
firmly joined with the lower part and cannot be
removed).
Fig. 5-1
Loosening fastening screws
Fig. 5-2
Removing the upper part of the
housing