Axiovert 40
B 46-0031 e 11/03
2Ć5
(6) Connection to the power outlet
CAUTION
Verify the operating voltage indicated on the rear panel of the instrument agrees with the
available line voltage.
•
Plug power cable (2Ć7ā/ā
4
) into the instrument socket and connect to the power outlet.
•
Switch on the instrument at the power switch (2Ć7ā/
ā3
) on the right of the instrument.
NOTE
When the instrument is switched off, the "0" marking is visible on the power switch.
•
The instrument is ready for operation when the microscope lamp is on.
•
In the event of a defect, check the two fuses (2Ć7ā/
ā2
):
for 100
…
240 V AC:
T 1 A/H, 5 x 20 mm (acc. to IEC 127)
For this purpose, remove the fuse holder (2Ć7ā/ā
1
) from the housing by pressing the two springs simultaneously
in the direction indicated by the arrows.
CAUTION
Only use the correct fuses.
Fig. 2Ć7
Connection to the power outlet
1
2
3
4
Содержание Axiovert 40
Страница 1: ...Operating Manual Axiovert 40 Inverted Microscope...
Страница 2: ......
Страница 3: ......
Страница 11: ...Axiovert 40 B 46 0031 e 11 03 VII GENERAL VIEW Axiovert 40 CFL...
Страница 12: ...Axiovert 40 11 03 B 46 0031 e VIII...
Страница 16: ...Axiovert 40 11 03 B 46 0031 e XII...
Страница 24: ...Axiovert 40 11 03 B 46 0031 e 1 8 1 4 System overview Fig 1 3 1 Axiovert 40 system overview...
Страница 25: ...Axiovert 40 B 46 0031 e 11 03 1 9 Fig 1 3 2 Axiovert 40 system overview...
Страница 26: ...Axiovert 40 11 03 B 46 0031 e 1 10 Fig 1 3 3 Axiovert 40 system overview...
Страница 65: ...Axiovert 40 B 46 0031 e 11 03 A 1 ANNEX Index List of abbreviations...
Страница 66: ...Axiovert 40 11 03 B 46 0031 e A 2...
Страница 72: ...Axiovert 40 11 03 B 46 0031 e A 8...