Axiovert 40
11/03 B 46-0031 e
2Ć8
(11) Switching on/setting the instrument
•
Unless already done, switch on the instrument at
the power switch (2Ć11ā/
ā5
).
•
After switching on, the lamp in the illuminator
(2Ć11ā/ā
4
) must light up.
•
Adjust the required brightness by turning the
brightness control (2Ć11ā/ā
6
).
•
If required, insert the appropriate attenuation filĆ
ter, green filter, conversion filter or white balance
filter (2Ć11ā/ā
3
) in the filter slider (2Ć11ā/ā
2
) and push
the slider into the light path.
•
Move slider Ph, H, Ph (2Ć11ā/ā
1
, slider Ph, H, Var
(2Ć4ā/
ā3
), slider Var, H, Var or slider Ph, H, PlasDIC
(2Ć4ā/
ā4
) to its mid-position.
NOTE
If no light appears in the viewing
path of the Axiovert 40 C or AxiovĆ
ert 40 CFL, actuate the "observaĆ
tion/camera" path selector switch
(2Ć11ā/
ā7
).
2.2
Start-up
When starting up the Axiovert 40 microscope for the first time, the instrument has to be unpacked in accorĆ
dance with Section 2.1.1, installed in accordance with Section 2.1.2, connected to the power outlet and
made ready for operation.
The Axiovert 40 microscope is supplied with factory-centered halogen bulb, which need not be recentered
even after you replaced the halogen bulb.
After any change of the lamp of the HBO 50 illuminator, however, you have to center this lamp (for that,
refer to Section 3.2 (3)).
Minor adjustment procedures, such as the adjustment of an annular diaphragm, are explained in the sections
covering the respective microscopic techniques.
Otherwise, settings are restricted to changing the objectives (swinging them into the light path) and the hanĆ
dling of LD objectives.
Fig. 2Ć11
Switching on/setting the instrument
1
6
2
3
4
7
5
Содержание Axiovert 40
Страница 1: ...Operating Manual Axiovert 40 Inverted Microscope...
Страница 2: ......
Страница 3: ......
Страница 11: ...Axiovert 40 B 46 0031 e 11 03 VII GENERAL VIEW Axiovert 40 CFL...
Страница 12: ...Axiovert 40 11 03 B 46 0031 e VIII...
Страница 16: ...Axiovert 40 11 03 B 46 0031 e XII...
Страница 24: ...Axiovert 40 11 03 B 46 0031 e 1 8 1 4 System overview Fig 1 3 1 Axiovert 40 system overview...
Страница 25: ...Axiovert 40 B 46 0031 e 11 03 1 9 Fig 1 3 2 Axiovert 40 system overview...
Страница 26: ...Axiovert 40 11 03 B 46 0031 e 1 10 Fig 1 3 3 Axiovert 40 system overview...
Страница 65: ...Axiovert 40 B 46 0031 e 11 03 A 1 ANNEX Index List of abbreviations...
Страница 66: ...Axiovert 40 11 03 B 46 0031 e A 2...
Страница 72: ...Axiovert 40 11 03 B 46 0031 e A 8...