
START-UP
Carl Zeiss
Installing optional components
Axio Examiner
54
M60-2-0003 e 05/2012
3.3.9
Replacing the filter in the
transmitted light filter wheel
•
Take off stage carrier, condenser, and
condenser carrier (see sections 3.1.9, 3.1.10
•
Take off the cover plate (Figure 3-41/
1
) from
the stand base.
•
Pull the filter to be replaced (Figure 3-41/
2
) up
and out of the corresponding position of the
filter wheel (Figure 3-41/
3
) using a pair of
tweezers.
•
Insert a new filter into the filter wheel
(Figure 3-41/
3
).
A filter position is always designed as a
permanently installed shutter position.
•
Repeat the procedure for more filter wheel
positions as required.
•
Insert the cover plate (Figure 3-41/
1
), and press
it down.
•
Attach stage carrier, condenser, and condenser
carrier (see sections 3.1.9, 3.1.10 and 3.1.12).
Figure 3-41 Replacing the filter in the
transmitted light filter wheel
Содержание Axio Examiner
Страница 1: ...Operating Manual Axio Examiner Research Microscope...
Страница 13: ...INTRODUCTION Axio Examiner List of interfaces Carl Zeiss M60 2 0003 e 05 2012 13 Figure 1 2 List of interfaces...
Страница 19: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 19 2 2 System overview...
Страница 20: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 20 M60 2 0003 e 05 2012...
Страница 21: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 21...
Страница 22: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 22 M60 2 0003 e 05 2012...
Страница 96: ......