START-UP
Carl Zeiss
Installing standard components
Axio Examiner
36
M60-2-0003 e 05/2012
3.1.12
Attaching, removing and adjusting
the height of a stage carrier
(1) Attaching the stage carrier
To attach the stage carrier at the stand, please
proceed as follows:
•
Hold the stage carrier (Figure 3-17/
3
) and the
mounted stage with your left hand, and insert it
at a slight angle first into the dovetail guide
(Figure 3-17/
1
) at the stand.
•
Hold the stage carrier as horizontally as
possible, and push it against the left-hand guide
surfaces. While doing so, swivel the stage
carrier to the right to ensure that the right-hand
side also engages in the dovetail guide at the
stand.
•
Tighten the clamping screw (Figure 3-17/
2
)
slightly with your right hand.
•
Align the stage carrier correspondingly at the
lower stop screw, and tighten the clamping
screw (Figure 3-17/
2
) without using excessive
pressure.
If the adapter is used for expanding the
specimen area (see section 3.3.1), align the
stage carrier at the upper stop screw.
•
Loosen the adjusting screws (Figure 3-17/
5
) at
the stage carrier supports using a ball head
screwdriver.
•
Push in the supports (Figure 3-17/
6
) and fasten
them at the stage carrier to the right and to the
left using screws (Figure 3-17/
7
).
•
Pull the supports down until the clamping rings
(Figure 3-17/
4
) rest on the stand.
•
Tighten the adjusting screws (Figure 3-17/
5
).
•
Tighten both clamping rings (Figure 3-17/
4
)
until you feel a slight resistance.
Do not overtighten the clamping rings.
Otherwise, the stage carrier will be pushed
up and/or bent.
Figure 3-17 Attaching the stage carrier
(stage not shown in the drawing)
Содержание Axio Examiner
Страница 1: ...Operating Manual Axio Examiner Research Microscope...
Страница 13: ...INTRODUCTION Axio Examiner List of interfaces Carl Zeiss M60 2 0003 e 05 2012 13 Figure 1 2 List of interfaces...
Страница 19: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 19 2 2 System overview...
Страница 20: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 20 M60 2 0003 e 05 2012...
Страница 21: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 21...
Страница 22: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 22 M60 2 0003 e 05 2012...
Страница 96: ......