INTRODUCTION
Axio Examiner
Equipment safety notices
Carl Zeiss
M60-2-0003 e 05/2012
9
Axio Examiner microscopes
including any original accessories may not be used unless for the microscopy
procedures as described in this operating manual.
The following instructions should be observed in particular:
The manufacturer will not assume any liability whatsoever for any other use even if such other
use only refers to single assemblies or parts. This shall also be applicable to any servicing or
repair work not performed by authorized service personnel. All guarantee or warranty claims
will be void in addition.
No operation of this equipment is allowed in a potentially explosive environment.
Do not operate this equipment unless positioned on a hard and nonflammable basis.
Do not insert the power plug into any connector receptacle not provided with an earth/ground
contact. Do not bypass the protective effect by using an extension cable without a protective
conductor.
Shut down the equipment and prevent it from being used inadvertently whenever a protective
measure is found not to be effective any longer. For restarting the equipment, call the Zeiss
customer service or the Carl Zeiss Microscopy Service.
The external power supply unit for 12 V DC 100 W for HAL 100 is designed for using line
voltage values within a range of between 100 and 240 V ±10 %, 50 / 60 Hz. There is no need
to switch over between different voltage values.
The power supply units for HBO 103 (ebq 100 dc) or XBO 75 (ebx 75 isolated) are designed for
a voltage range of between 100 and 240 V AC, with 50 to 60 Hz. These units automatically
adapt to the line voltage applied. Accordingly, the voltage does not need to be switched over
for these devices either.
Before you switch on the equipment, check whether it is suitable for the line voltage available.
Unplug the power plug before opening the equipment and before replacing any fuse. Do not
use any fuse unless it is suitable for the rated current provided. Do not use any makeshift fuse,
and do not short-circuit the fuse holders.
Axio Examiner microscopes are not equipped with any particular device for protection against
caustic, potentially infectious, toxic, radioactive or other specimens which might impair health.
Please observe all statutory requirements, and especially all national regulations for accident
prevention whenever working with such specimens.
Содержание Axio Examiner
Страница 1: ...Operating Manual Axio Examiner Research Microscope...
Страница 13: ...INTRODUCTION Axio Examiner List of interfaces Carl Zeiss M60 2 0003 e 05 2012 13 Figure 1 2 List of interfaces...
Страница 19: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 19 2 2 System overview...
Страница 20: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 20 M60 2 0003 e 05 2012...
Страница 21: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 21...
Страница 22: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 22 M60 2 0003 e 05 2012...
Страница 96: ......