Mechanical cleaning/disinfection (validated process)
Phase
Step
T[˚C/ ˚F]
t[min]
Waterquality
Comment
I
Pre-rinse
< 40/104(cold)
3
T-W
Emptying
II
Cleaning
55/131
10
T-W
neodisher®
MediClean Dental
Emptying
III
Rinse
< 40/104
2
T-W
Emptying
IV
Rinse
< 30/86(cold)
1
VE-W
Emptying
V
Thermo-
disinfection
90/194
5
VE-W
Emptying
VI
Drying
> 90/104
30
Seite 39/69
Page 39/47
JMA-Optic
T-W: Drinking water
VE-W: Fully demineralized water (demineralized, microbiologically
at least drinking water quality)
Validated with: neodisher® MediClean Dental in a Miele PG 8535
Workflow:
■
Carefully place the product in a suitable rinsing basket.Make sure that the product does
not touch any other instruments/instrument parts (no rinsing shadow).
Start a tested program with the following characteristics:
■
Sufficient cleaning effect.
■
Final rinse with distilled or fully demineralized water.
■
Thermal disinfection with deionized water: 5 minutes at 90 - 95 C , A0 >=3000.
■
Sufficient product drying (no visible liquid residues).
■
Check and pack the components immediately after removal,
8.4./8.5.Inspection and packaging, p. 39/40
.
8.4. Inspection
All components must be checked for the following in each case after cleaning/disinfection:
■
Damage to the product, e.g. cracks, chipping.
■
Damage to the surface
■
Deformations on the product or parts of the product.
If the product still has visible contamination after the end of the program, repeat the pre-cleaning
and machine cleaning/disinfection. Damaged components must not be used any further.