IOMOD10F Calibration Module for Z-Wave 500 Series IC and modules - User Manual - V2.06 06/10/2017
20
The illustration below shows how to connect an external power supply to the
IOMOD10
module via a
2.1 mm centre positive jack socket.
Please note:
A suitable 3.3V DC regulated @ 1.8A power supply for use with this module is supplied as standard
by Equinox with the
IOMOD10F
module kit. Please use the Equinox power supply as this has been
checked for use with the
IOMOD10F
module.
1.3 Setting up the IOMOD10F module ready for production
The
‘IOMOD10F - Z-Wave calibration module’
is designed to be used with an Equinox
‘ISPnano
Series 4’
programmer.
Instructions:
•
Plug the module into an Equinox
‘ISPnano Series 4’
programmer
•
Connect the
'Z-Wave Target System (DUT)'
to either the
'Quick connect'
connector (via
wires) or to the
10-way IDC connector
via a 10-way ribbon cable.
•
Manually connect any power and GROUND signals required between the programmer and the
Target Board (DUT)
•
Set the '
Calibration Clock Output Pin Select'
(LK2) configuration jumper to select which pin
the
'Calibration frequency'
is output on.
•
Run a
'Calibration script'
using the Equinox
ISP-PRO
software utility in order to perform the
'XTAL calibration'
function.