1. Indications for use

The YH-550 self-adjusting device and YH-350 are indicated for the treatment of Obstructive Sleep 
Apnea (OSA) in patients weighing more than 66 lb (30 kg). They are intended for home use only.
YH-550 and YH-350 consist of main device, water tank and power adapter. Mask and air tube are 
purchased. Mask and air tube are type BF applied part. The patient is an intended operator.

2. Contraindications

Positive airway pressure therapy may be contraindicated in some patients with the following pre-
existing conditions:

    

Severe bullous lung disease, pneumothorax, pathologically low blood pressure, dehydration, 

cerebrospinal fluid leak, recent cranial surgery, or trauma

3. Warnings

 

y

Do not maintain the device while it is in use, otherwise it may lead to unacceptable risk.

 

y

Do not modify this equipment without authorization of the manufacturer or open this device by 
yourself, contact YUWELL when the device need to be repaired.

 

y

This device cannot be used for life support. It may be shut down by turning off the electricity, but 
no unacceptable risk will happen. 

 

y

The device can't be used in the environment where the air is mixed with flammable anesthetic 
gases or nitrous oxide gases. (Non-AP and Non-APG series PAP)

 

y

The sources of oxygen must be located more than 1 m from the BreathCare PAP to avoid the risk 
of fire and burns.

 

y

Supplemental oxygen must not be used while smoking or in the presence of an open flame.

 

y

Always make sure that the device is turned on and airflow generated before the oxygen supply 
is turned on. Always turn the oxygen supply off before the device is turned off, so that unused 
oxygen will not accumulate within the device enclosure or create a risk of fire.

 

y

Make sure that you arrange the cables and air tube so that they will not twist around the head or 
neck. Otherwise it may result in strangulation.

 

y

If you notice any unexplained change in the performance of the device, if it is making unusual 
sounds, if the device or the power supply are dropped or mishandled, or if the enclosure is broken, 
discontinue use and power off the equipment, then contact your provider or YUWELL Service 
Center.

 

y

Do not place the device where it can be bumped or where someone is likely to trip over the power 
cord.

 

y

Do not block the air tube and/or air inlet of the device while in operation otherwise it could cause 
overheating of the device.

 

y

Keep the area around the device dry, clean and clear of anything (e.g. clothes, bedding, lint, 
dust or direct sunlight) that could block the air inlet, cover the power supply unit, affect patients’ 
breathing or reduce the life of device.

 

y

Place the device on the stable table. Placing device on the soft, out of flatness surface is forbidden. 

 

y

Keep device far away from water.

 

y

Make sure the power cord and plug are in good condition and the equipment is not damaged.

 

y

Keep the power cord away from hot surfaces.

 

y

Beware of electrocution. Do not immerse the device, power supply or power cord in water. If liquids 
are spilled into or onto the device. Unplug the device and let the parts dry naturally and contact 
YUWELL Service Center. 

 

y

Always unplug the device before cleaning and make sure that all parts are dry before plugging it 
back in. 

 

y

This device cannot be used for multiple patients.

 

y

Use of this equipment adjacent to or stacked with other equipment should be avoided because 
it  could  result  in  improper  operation.  If  such  use  is  necessary,  this  equipment  and  the  other 
equipment should be observed to verify that they are operating normally. 

 

y

Use  of  accessories,  transducers  and  cables  other  than  those  specified  or  provided  by  the 
manufacturer of this equipment could result in increased electromagnetic emissions or decreased 
electromagnetic immunity of this equipment and result in improper operation.

 

y

Portable  RF  communications  equipment  (including  peripherals  such  as  antenna  cables  and 

Содержание YH-350

Страница 1: ...nology Co Ltd Add of manufacturer No 9 Jinfeng Road Suzhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 Web address www yuyue com cn Shanghai International Holding Corp GmbH...

Страница 2: ...9 14 Traveling 20 15 Trouble shooting 20 16 Specified of technology 23 17 Symbols 27 18 Limited warranty 28 19 Repairing 29 20 List of cables 29 21 Technical description 29 22 Warranty card 34 Content...

Страница 3: ...ngulation y y If you notice any unexplained change in the performance of the device if it is making unusual sounds if the device or the power supply are dropped or mishandled or if the enclosure is br...

Страница 4: ...es with device Non YUWELL parts may reduce effectiveness of the treatment and or damage the device y y Use only vented masks recommended by YUWELL or by your doctor with this device Fitting the mask w...

Страница 5: ...er tank 1 Air filter 2 Air tube 1 SD card 1 Power adapter and Cable as part of the main device 1 User manual 1 Bag 1 7 Picture and explanation for product Air outlet Dial START STOP button Interface s...

Страница 6: ...fficult to operate the disconnection device The following section will help you install your device by yourself 1 2 1 Place the device on a stable level surface Rotate the knob cover component from th...

Страница 7: ...the device is delivering gas pressure the heating plate is heating including the situations that the screen shows Large Leakage and the backlight of the screen is off When the level of the humidity is...

Страница 8: ...rn up the humidity If you are getting any moisture in your mask turn down the humidity You can set Humidity level between 0 to 6 where 0 means this function is disabled 1 is the lowest humidity settin...

Страница 9: ...ur change y y Highlight the Back press dial to back to home page Set Pressure y y At clinical page choose CPAP mode Highlight Initial Pressure or Therapy Pressure press and rotate dial to set suitable...

Страница 10: ...0 4 0 Initial Pressure Picture 6 1 Clinical menu CPAP Picture 6 2 Clinical menu APAP 12 Caring for your device It is important that you regularly clean your device to make sure you receive optimal th...

Страница 11: ...tank and air tube thoroughly and allow to dry out of direct sunlight and or heat Temperature should not exceed 40 Warning It is generally not necessary to sterilize your water tank if you are followin...

Страница 12: ...tube firmly onto the assembled mask 13 Therapy data BreathCare PAP records your therapy data for you and your care provider so they can view and make changes to your therapy if required The data is re...

Страница 13: ...ck After therapy start the screen backlight went off In other case the power do not connect firmly Press dial to turn on the screen light Check the connection of power ensure it connect with device fi...

Страница 14: ...el Sound pressure level measured according to ISO 80601 2 70 2015 CPAP mode BreathCare PAP 32dB A Sound power level Sound power level measured according to ISO 80601 2 70 2015 CPAP mode BreathCare PAP...

Страница 15: ...rmance Humidification system output 12mg L Relative humidity 50 Humidification system Pressure drop ISO 5367 2014 Flowrate L min Pressure drop cmH2O 30 0 18 60 0 6 90 1 2 Gas leakage at the maximum op...

Страница 16: ...an Authorized Representative Environmental information EU directive 2012 19 EE Waste Electrical and Electronic Equipment WEEE Protected against finger sized objects and against dripping water on the s...

Страница 17: ...ssions CISPR 11 Class B Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC61000 3 3 Complies f f 21 2 Compliance information for Immunity test Immunity test Compliance...

Страница 18: ...The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alte...

Страница 19: ...ce organizations that have not been expressly authorized by Yuwell to perform such repairs Any damage caused as accident act of god or human factor Product which is not involved in quality warranty sh...

Страница 20: ...damage caused as a result of improper use abuse modification or alteration of the product Repairs carried out by any service organizations that have not been expressly authorized by Yuwell to perform...

Отзывы: