and 3 is the highest level of FPS. The higher FPS level means more comfortable expiration you will 
get.

 

y

Date

: Adjust the date (year, month, day) displayed on the device.

 

y

Time

: Adjust the time (hour, minute) displayed on the device.

 

y

Language

: You can choose English or Chinese.

 

y

Filter Cotton Reminder

: When this function is enabled, this can remind you to check and change 

the filter.

 

y

Reset

: When this function is enabled, parameters of the device will be reset to the factory defaults 

and data will be erased.

Info

Humidity

Setup

Ramp

Sound Reminder
Smart Start/Stop

OFF

FPS Level

0

Date

2019-5-4

OFF

         

                                    Picture 5-1: Home                                  Picture 5-2: Setup

Date

2019-5-4

Time

9:27

Language

EN

OFF

Filter Cotton Reminder

Language

EN

Filter Cotton Reminder OFF
Reset
Back<<

>>

         

                                    Picture5-3: Setup                                     Picture 5-4: Setup

 

f

11.2  Function that patient must use under the direction of doctor (Clinical menu)

The patient should use the device by the direction of physician.

① Set Mode (CPAP or APAP)

 

y

At home page, press 

 and dial at the same time, the clinical menu will be shown on screen.

 

y

Rotate the dial, highlight the "Mode" , and then select CPAP or APAP.

 

y

After mode setting, press dial to save your change.

 

y

Highlight the "Back", press dial to back to home page.

② Set Pressure 

 

y

At clinical page, choose "CPAP" mode, Highlight "Initial Pressure" or "Therapy Pressure" , press 
and rotate dial to set suitable value. 

 

y

You can rotate the dial to the right or to the left to rise or decrease therapy pressure (each step is 0.5 
cmH

2

O / hPa).

 

y

After setting therapy pressure, press dial to save your change.

 

y

Highlight the "Back", press dial to back to home page.

 

y

You can set "Max Pressure" , "Min Pressure" , "Initial Pressure" and "Therapy Pressure" of APAP 
mode according to above method.

Notice

When you try to make the max pressure lower than min pressure,initial pressure or therapy pressure, 
these pressures will be equal to the max pressure.

③ Set Pressure Unit

 

y

At clinical page, highlight "Unit".

 

y

Rotate the dial to the right or left, set your pressure unit (cmH

2

O or hPa).

 

y

After setting pressure unit, highlight the "Back" , press dial to back to home page.

④ Set FPS Level

 

y

At clinical page, Highlight "FPS Level".

 

y

Rotate the dial to the right or left, set your level of FPS (0-3 level).

 

y

After setting level of FPS, highlight the "Back", press dial to back to home page.

⑤ Set Ramp

 

y

At clinical page, Highlight "Ramp".

 

y

Rotate the dial to the right or left, set your ramp time (0-45minutes, step value is 5 minutes).

 

y

After setting Ramp, highlight the "Back", press dial to back to home page.

Содержание YH-350

Страница 1: ...nology Co Ltd Add of manufacturer No 9 Jinfeng Road Suzhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 Web address www yuyue com cn Shanghai International Holding Corp GmbH...

Страница 2: ...9 14 Traveling 20 15 Trouble shooting 20 16 Specified of technology 23 17 Symbols 27 18 Limited warranty 28 19 Repairing 29 20 List of cables 29 21 Technical description 29 22 Warranty card 34 Content...

Страница 3: ...ngulation y y If you notice any unexplained change in the performance of the device if it is making unusual sounds if the device or the power supply are dropped or mishandled or if the enclosure is br...

Страница 4: ...es with device Non YUWELL parts may reduce effectiveness of the treatment and or damage the device y y Use only vented masks recommended by YUWELL or by your doctor with this device Fitting the mask w...

Страница 5: ...er tank 1 Air filter 2 Air tube 1 SD card 1 Power adapter and Cable as part of the main device 1 User manual 1 Bag 1 7 Picture and explanation for product Air outlet Dial START STOP button Interface s...

Страница 6: ...fficult to operate the disconnection device The following section will help you install your device by yourself 1 2 1 Place the device on a stable level surface Rotate the knob cover component from th...

Страница 7: ...the device is delivering gas pressure the heating plate is heating including the situations that the screen shows Large Leakage and the backlight of the screen is off When the level of the humidity is...

Страница 8: ...rn up the humidity If you are getting any moisture in your mask turn down the humidity You can set Humidity level between 0 to 6 where 0 means this function is disabled 1 is the lowest humidity settin...

Страница 9: ...ur change y y Highlight the Back press dial to back to home page Set Pressure y y At clinical page choose CPAP mode Highlight Initial Pressure or Therapy Pressure press and rotate dial to set suitable...

Страница 10: ...0 4 0 Initial Pressure Picture 6 1 Clinical menu CPAP Picture 6 2 Clinical menu APAP 12 Caring for your device It is important that you regularly clean your device to make sure you receive optimal th...

Страница 11: ...tank and air tube thoroughly and allow to dry out of direct sunlight and or heat Temperature should not exceed 40 Warning It is generally not necessary to sterilize your water tank if you are followin...

Страница 12: ...tube firmly onto the assembled mask 13 Therapy data BreathCare PAP records your therapy data for you and your care provider so they can view and make changes to your therapy if required The data is re...

Страница 13: ...ck After therapy start the screen backlight went off In other case the power do not connect firmly Press dial to turn on the screen light Check the connection of power ensure it connect with device fi...

Страница 14: ...el Sound pressure level measured according to ISO 80601 2 70 2015 CPAP mode BreathCare PAP 32dB A Sound power level Sound power level measured according to ISO 80601 2 70 2015 CPAP mode BreathCare PAP...

Страница 15: ...rmance Humidification system output 12mg L Relative humidity 50 Humidification system Pressure drop ISO 5367 2014 Flowrate L min Pressure drop cmH2O 30 0 18 60 0 6 90 1 2 Gas leakage at the maximum op...

Страница 16: ...an Authorized Representative Environmental information EU directive 2012 19 EE Waste Electrical and Electronic Equipment WEEE Protected against finger sized objects and against dripping water on the s...

Страница 17: ...ssions CISPR 11 Class B Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC61000 3 3 Complies f f 21 2 Compliance information for Immunity test Immunity test Compliance...

Страница 18: ...The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alte...

Страница 19: ...ce organizations that have not been expressly authorized by Yuwell to perform such repairs Any damage caused as accident act of god or human factor Product which is not involved in quality warranty sh...

Страница 20: ...damage caused as a result of improper use abuse modification or alteration of the product Repairs carried out by any service organizations that have not been expressly authorized by Yuwell to perform...

Отзывы: