Yuwell W10 Скачать руководство пользователя страница 4

03

04

It will affect the measurement accuracy if the wrist circumference is out 
of the given value.
Do  not  measure continuously. (2~3    minutes or  more  should  be  spared 
between two measurements).
Do not use mobile phone near the monitor.
Do not keep the cuff in the aerated state for a long time.
The patient is an intended operator, this monitor is used for adult whom 
more than 12 years old.
Do not swallowed small parts that may cause choking hazard.
The device must not be used with high frequency surgical equipments.
Do not use the CUFF over a wound arm or being on an intravenous drip. 
Consult your doctor if using the device on the arm with an arteriovenous 
(A-V) shunt.
Do not use the cuff on the arm where the side of a mastectomy or lymph 
node clearance.
Note  that  It  will  lose  function  of  other  device  on  the  same  limb  while 
cuff inflating.
Do not allowed frequent measurements, the resulting restriction of the 
blood flow may cause injury.
Please using soft dry cloth stained with 75% ethanol to clean the device 
in the case of many people use it,but do not let the ethanol flow into the 
monitor and cuff.
Please report serious incident that has occurred in relation to the device 
to the manufacturer and the competent authority of the Member State.

Important  notes

m

2.  Adapter
    

1. There is no adapter in the accessories of this product, the user can 

  

     it using the complimentary USB cable.

2. If you need to buy the yuwell special adapter that matches this product, 

please contact the local dealer for consultation.

charge

adapter

USB  cable

Tips!

Please turn off the device and unplug the adapter before cleaning.
The AC adapter cable may cause accidental strangulation in infants.
Please pay attention to product storage to prevent damage caused by pets, 
pests or children.
Operator  can't  touch  the  accessible  part  of  adapter/main  unit  and  patient 
at the same time.
According to the local laws and regulations to deal with battery.
Don't  interconnect  this  device  to  other  equipment  not  described  in  the 
user's manual.
Note:  Do  not  diagnose  with  the  measurement,  Please  follow  doctor's 
instruction.

Statement: 

If the monitor has not been stored in the required temperature 

and humidity range, it may not conformance to specification.

Important  notes

Please take attention that changes or modification not expressly approved 
by the party responsible for compliance could void the user's authority to 
operate the equipment.
This device complies with Part 15 of the FCC Rules. Operation is subject to 
the following two conditions: 
(1) This device may not cause harmful interference, and 
(2) This device must accept any interference received, including 
interference that may cause undesired operation.

Tips!

Содержание W10

Страница 1: ...Monitor Please read the user s manual closely before using SYS mmHg mmHg DIA P min m Shanghai International Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany Date of 2021 05 issue JIANGSU...

Страница 2: ...tents Cautions 01 Important notes 02 05 Parts 06 Product charging 08 10 Voice and unit setting 11 Using method of cuff 12 Measuring posture 13 Start measuring 15 Memory function 19 Auxiliary function...

Страница 3: ...ing during transportation or transportation 01 This product is intended to measure the blood pressure and pulse rate of adult whom more than 12 years old at household or medical center not suitable fo...

Страница 4: ...f Please report serious incident that has occurred in relation to the device to the manufacturer and the competent authority of the Member State Important notes Please turn off the device and unplug t...

Страница 5: ...essary Do not use detergent to clean Do not let water seep into the device Manufacture will make available on request circuit diagrams component part lists descriptions calibration instructions or oth...

Страница 6: ...lease unplug the power adapter from the socket after charging icon flashs during charging and the The showing 25 or less mean the battery is and needs to be charged The icon appearing and showing 5 or...

Страница 7: ...ate as following steps For example setting the date as 2020 8 11 and and time as 08 18 Time and date setting Product charging Do not place equipment with lithium batteries near the fire source Do not...

Страница 8: ...easuring unit setting Press the button to choose unit between mmHg and kPa Press to finish setting M D M D Using method of cuff 3 Please stick the cuff firmly or it may get loose while measurement and...

Страница 9: ...another hand 3 Do not hold the arm hard while measurement If it is necessary please support the arm with a soft pillow Figure 2 The correct wrist posture as Figure 2 NOTE Relax fingers and keep a natu...

Страница 10: ...rements should be at least 2 3 minutes or longer SYS mmHg mmHg DIA P min m Sit still for 5 minutes before the measurement Full screen display state 1 Measurement All icons on screen appear for 1 sec a...

Страница 11: ...lay kPa display Press the button to shut down Then it will power off automatically in 3 minutes without any operation 3 Take off the cuff 4 Shut down Please contact with doctor for advice if the SYS i...

Страница 12: ...mHg DIA P min m m SYS mmHg mmHg DIA P min M D M D The 74th data may be replaced by the 73rd when the memory capacity is full Tips Press the both button and m button in memory display until the display...

Страница 13: ...onal personnel to enter the static mode to test the monitor through standard pressure gauge Warning normal users don t need to know this function and also do not operate The company will not take any...

Страница 14: ...ion will accelerate atherosclerosis and cause nerve ending dysaemia The blood pressure on wrist and upper arm may have a significant difference on these patients Even if in unsuitable conditions healt...

Страница 15: ...t under control Do not smoke exercise regularly reduce salt and fat intake have regular 3 Why measure Blood Pressure at home Blood pressure measured at a clinic or doctor s office may cause apprehensi...

Страница 16: ...al blood pressure Regular readings give a more comprehensive blood pressure history Be sure to note date and time when recording your blood pressure Consult doctor to interpret the blood pressure data...

Страница 17: ...a clinic or doctor s office Pump works but the pressure doesn t rise Wrong indication Err 4 Err 5 Err 6 Err 7 The table below shows the possible fault displaying icon possible reason and solutions Un...

Страница 18: ...1000mA Maximum battery life the battery can be used about 250 times on a full charge Pressure Diastolic 40 210mmHg Systolic 60 260mmHg Suitable wrist circumference 13 5 19 5cm Dimensions Approx L89 m...

Страница 19: ...ts or instruments please inspect and verify if the instrument could run normally Electromagnetic compatibility information Guidance and manufacture s declaration electromagnetic emission Emission test...

Страница 20: ...2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters suc...

Страница 21: ...20 1700 1990 GSM 1800 TETRA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS 0 3 28 1970 Pulse modulationb 217 Hz 2 870 1845 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 0 3 2450 Pulse modulation b 217 H...

Отзывы: