Electromagnetic compatibility information
Electromagnetic compatibility information
Table 2
Immunity
test
IEC 60601
test level
Compliance level
30A/m
50Hz or 60Hz
3A/m
50Hz or 60Hz
Guidance and manufacturere' s declaration-electromagnetic immunity
Electrostatic
discharge(ESD)
IEC 61000-4-2
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
±8 kV
contact
±15
kV air
±8 kV
contact
±15
kV air
Guidance and manufacture s declaration-electromagnetic emission
'
Immunity test
IEC 60601 test level
Compliance level
Electrostatic
discharge
(ESD) IEC
61000-4-2
Electrical
fasttransient
/burst IEC
61000-4-4
±2 kV
100kHz repetition
frequency
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
±2 kV
100kHz repetition
frequency
Surge IEC
61000-4-5
±1 kV line(s)to line(s)
±2 kV line(s)to earth
±1 kV line(s)to line(s)
0% U ; 0.5 cycle
T
o
o
o
o
o
o
At 0 , 45 , 90 , 135 , 180 , 225 ,
o
o
270 and 315
Voltage dips
IEC 61000-4-11
NOTE: UT is the a.c. mains voltage prior to application of the test level.
30A/m
50Hz or 60Hz
Power frequency
(50/60 Hz)
magnetic
field IEC
61000-4-8
0% U ; 0.5 cycle
T
o
o
o
o
o
o
At 0 , 45 , 90 , 135 , 180 , 225 ,
o
o
270 and 315
30A/m
50Hz or 60Hz
The
is intended for use in the
electromagnetic environment specified below. The customer or the user of the
should assure that it is used in
such an environment.
W10/W10S/W10T/W10F
W10/W10S/W10T/W10F
Electronic Blood Pressure Monitor
Electronic Blood Pressure Monitor
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture s declaration-electromagnetic immunity
'
Immunity test
IEC 60601 test level
Compliance level
3 Vrms
0,15 MHz-80 MHz
6 V rms in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % Am at 1 kHz
Conducted RF
IEC 61000-4-6
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field
strengths
from
fixed
transmitters
,
such
as
base
stations
for radio
(
cellular
/
cordless
)
telephones
and
land
mobile
radios
,
amateur radio
,
AM and
FM radio broadcast and
TV
broadcast
cannot
be predicted
theoretically
with
accuracy
.
To
assess the electromagnetic environment
due
to fixed
RF
transmitters
,
an
electromagnetic
site survey should
be
considered
.
If
the
measured
field
strength
in
the location
in
which
the
is used exceeds
the
applicable
RF
W10/W10S/W10T/W10F
compliance
level
above
,
the
should
be
observed
to
verify normal
operation
.
W10/W10S/W10T/W10F
If
abnormal
performance
is
observed
,
additional measures
may
be
necessary
,
such
as
reorienting
or
relocating
the
.
W10/W10S/W10T/W10F
b
Over
the
frequency
range
150
kHz
to
80
MHz
,
field
strengths should
be
less
than
3
V
/
m
.
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to
2.7 GHz
3 Vrms
0,15 MHz-80 MHz
6 V rms in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % Am at 1 kHz
10 V/m
80 MHz to
2.7 GHz
The
is intended for use in the
electromagnetic environment specified below. The customer or the user of the
W10/W10S/W10T/W10F
should assure that it is used in
such an environment.
W10/W10S/W10T/W10F
Electronic Blood Pressure Monitor
Electronic Blood Pressure Monitor
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
0% U ; 1 cycle and 70% U ;
T
T
o
25/30 cycles Single phase: at 0
0% U ; 250/300 cycles
T
Voltage interruptions
IEC 61000-4-11
0% UT; 1 cycle and 70% UT;
o
25/30 cycles Single phase: at 0
0% U ; 250/300 cycles
T