Features and technical parameters

Features and technical parameters

1. Features 

Small delicate design
74 groups of memory
Average value display of three times measurement 
kPa & mmHg units display
Automatically shut down in 3 minutes after measurement.

2. Technical parameters

Displaying: LCD digital display
Measuring method: oscillation mensuration

Measuring range

0~300mmHg

Cuff Pressure: 0-450mmHg

     

Pulse rate: 40~200 beats/min

Precision

Pressure: within ±3mmHg(±0.4kPa)
Pulse rate: within 5% of reading value
Working system: continuous operation
IP Classification: IP22
 

                                                                                   

Electric classification: Class II, type BF application part
Service life: 5 years or 50 thousand times

Power supply

battery:DC 3.7V
AC adapter:input 100-240V~        50/60Hz  0.35A MAX

                             

output 

 

5V     1000mA

Maximum battery life: the battery can be used about 250 times on a full charge

Pressure: 
Diastolic: 40-210mmHg       
Systolic: 60-260mmHg

Suitable wrist circumference: 13.5-19.5cm
Dimensions:Approx.L89 mm x W62 mm x H21 mm
Wight: about 109g

4. Operation and storage conditions:

Operating temperature/humidity: +5℃~+40℃/15%~90%RH(no-condensation)
Storage temperature/humidity: -20℃~ +55℃/15%~90%RH(no-condensation)
Operating and storage atmospheric pressure: 70kPa ~ 106kPa

Operation environment:

Avoid electromagnetic interference, violent shock and noise environment.

The contact materials detail of product

Part

Rear Cover

ABS
ABS

Magic paster
cloth

PC

Dacron

Panel

The  SPHYGMOMANOMETER  was  clinically  investigated  according  to  the 
requirement  of

 

 ISO 81060-2.

The  SPHYGMOMANOMETER  complies  with  IEC 80601-2-30.

Statement!

Nylon

Top Cover

Cuff

Material

against ingress of solid foreign objects: ≥12.5mm diamater

against ingress of water with harmful effects: dripping (15°tilted)

Содержание W10

Страница 1: ...Monitor Please read the user s manual closely before using SYS mmHg mmHg DIA P min m Shanghai International Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany Date of 2021 05 issue JIANGSU...

Страница 2: ...tents Cautions 01 Important notes 02 05 Parts 06 Product charging 08 10 Voice and unit setting 11 Using method of cuff 12 Measuring posture 13 Start measuring 15 Memory function 19 Auxiliary function...

Страница 3: ...ing during transportation or transportation 01 This product is intended to measure the blood pressure and pulse rate of adult whom more than 12 years old at household or medical center not suitable fo...

Страница 4: ...f Please report serious incident that has occurred in relation to the device to the manufacturer and the competent authority of the Member State Important notes Please turn off the device and unplug t...

Страница 5: ...essary Do not use detergent to clean Do not let water seep into the device Manufacture will make available on request circuit diagrams component part lists descriptions calibration instructions or oth...

Страница 6: ...lease unplug the power adapter from the socket after charging icon flashs during charging and the The showing 25 or less mean the battery is and needs to be charged The icon appearing and showing 5 or...

Страница 7: ...ate as following steps For example setting the date as 2020 8 11 and and time as 08 18 Time and date setting Product charging Do not place equipment with lithium batteries near the fire source Do not...

Страница 8: ...easuring unit setting Press the button to choose unit between mmHg and kPa Press to finish setting M D M D Using method of cuff 3 Please stick the cuff firmly or it may get loose while measurement and...

Страница 9: ...another hand 3 Do not hold the arm hard while measurement If it is necessary please support the arm with a soft pillow Figure 2 The correct wrist posture as Figure 2 NOTE Relax fingers and keep a natu...

Страница 10: ...rements should be at least 2 3 minutes or longer SYS mmHg mmHg DIA P min m Sit still for 5 minutes before the measurement Full screen display state 1 Measurement All icons on screen appear for 1 sec a...

Страница 11: ...lay kPa display Press the button to shut down Then it will power off automatically in 3 minutes without any operation 3 Take off the cuff 4 Shut down Please contact with doctor for advice if the SYS i...

Страница 12: ...mHg DIA P min m m SYS mmHg mmHg DIA P min M D M D The 74th data may be replaced by the 73rd when the memory capacity is full Tips Press the both button and m button in memory display until the display...

Страница 13: ...onal personnel to enter the static mode to test the monitor through standard pressure gauge Warning normal users don t need to know this function and also do not operate The company will not take any...

Страница 14: ...ion will accelerate atherosclerosis and cause nerve ending dysaemia The blood pressure on wrist and upper arm may have a significant difference on these patients Even if in unsuitable conditions healt...

Страница 15: ...t under control Do not smoke exercise regularly reduce salt and fat intake have regular 3 Why measure Blood Pressure at home Blood pressure measured at a clinic or doctor s office may cause apprehensi...

Страница 16: ...al blood pressure Regular readings give a more comprehensive blood pressure history Be sure to note date and time when recording your blood pressure Consult doctor to interpret the blood pressure data...

Страница 17: ...a clinic or doctor s office Pump works but the pressure doesn t rise Wrong indication Err 4 Err 5 Err 6 Err 7 The table below shows the possible fault displaying icon possible reason and solutions Un...

Страница 18: ...1000mA Maximum battery life the battery can be used about 250 times on a full charge Pressure Diastolic 40 210mmHg Systolic 60 260mmHg Suitable wrist circumference 13 5 19 5cm Dimensions Approx L89 m...

Страница 19: ...ts or instruments please inspect and verify if the instrument could run normally Electromagnetic compatibility information Guidance and manufacture s declaration electromagnetic emission Emission test...

Страница 20: ...2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters suc...

Страница 21: ...20 1700 1990 GSM 1800 TETRA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS 0 3 28 1970 Pulse modulationb 217 Hz 2 870 1845 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 0 3 2450 Pulse modulation b 217 H...

Отзывы: