Phenomenon
Basic EMC standard
or test method
Professional
healthcare
facility environment
Home healthcare
facility environment
Electrostatic
discharge
IEC 61000-4-2
+/- 8 kV contact
+/- 2 kV, +/- 4 kV, +/- 8 kV, +/- 15 kV air
Radiated RF EM
fields
IEC 61000-4-3
3V/m
80MHz-2.7GHz
80%AM at 1kHz or 2Hz
10V/m
80MHz-2.7GHz
80%AM at 1kHz or
2Hz
Proximity
fields
from RF
wireless
communications
equipment
IEC 61000-4-3
See the RF wireless communication equipment
table in "Recommended minimum separation
distances".
Rated
power
frequency
magnetic fields
IEC 61000-4-8
30A/m; 50 Hz or 60Hz
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
equipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
Immunity
Test Level
(V/m)
385
380-390 TETRA 400
Pulse
modulation
b)
18Hz
1.8
0.3
27
450
430-470
GMRS 460,
FRS 460
FM
c)
±5kHz
deviation
1kHz sine
2
0.3
28
710
704-787
LTE Band
13, 17
Pulse
modulation
b)
217Hz
0.2
0.3
9
745
780
810
800-960
GSM800/90
0, TETRA
800, iDEN
820, CDMA
850, LTE
Band 5
Pulse
modulation
b)
18Hz
2
0.3
28
870
930
1720
1700-
GSM1800;
Pulse
2
0.3
28