![Yolanda TH10A Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/yolanda/th10a/th10a_user-manual_3440436019.webp)
page
19 of
- Do not use 12Y-up mode to measure the body temperature of people aged 12 and below,
Do not use 12Y-down mode to measure the body temperature of people over 12 years old,
otherwise the measurement results may be slightly biased.
Notes:
- Before measuring forehead, make sure there is no hair, sweat, cosmetic,
perspiration
or hat
on it.
- It is normal that there may be temperature difference depending on various skin types, since
different skin type will reflect different voltage of infrared ray.
- Avoid any drafts (e.g. form nasal specs, air conditioning
…
)
9.3 Temperature hint
Measurement mode
Measuring temperature
Backlight
Buzzer sound
LCD display
Ear mode
/Forehead
mode
12Y-up
mode
< 32.0
℃
(89.6
℉
)
Red
bi---
Lo
32.0
℃
(89.6
℉
) ~ 37.7
℃
(99.9
℉
)
Cyan
bi---
Measurement
value
37.8
℃
(100.0
℉
)~39.4
℃
(102.9
℉
)
Yellow
bi-bi-bi-bi
39.5
℃
(103.1
℉
)~43.0
℃
(109.4
℉
)
Red
bi-bi-bi-bi
>
43.0
℃
(109.4
℉
)
Red
bi---
Hi
12Y-down
mode
< 32.0
℃
(89.6
℉
)
Red
bi---
Lo
32.0
℃
(89.6
℉
) ~ 37.6
℃
(99.7
℉
)
Cyan
bi---
Measurement
value
37.7
℃
(99.9
℉
)~38.5
℃
(101.3
℉
)
Yellow
bi-bi-bi-bi
38.6
℃
(101.5
℉
)~43.0
℃
(109.4
℉
)
Red
bi-bi-bi-bi
>
43.0
℃
(109.4
℉
)
Red
bi---
Hi
Operation environment temperature < 15
℃
(59.0
℉
)
Cyan
bi---
Er1
Operation environment temperature
>
40
℃
(104.0
℉
)
Cyan
bi---
Er2
Note: The buzzer function in the table above is built when the buzzer sound is turned on.
9.4 Voice adjustment
On the power on status, press the “
”
button for 2s,the buzzer sound will be turned off
,
there
will be no buzzer sound during subsequent operations. Pressing the“
”
button for 2s again,
the voice function will be turned on with a “bi”sound.
9.5 Memory mode