If Essential Performance is lost or degraded due to electromagnetic disturbances, this may result
in inaccurate measurement and lead to mislead patients , please read below important
information before to avoid possible electromagnetic disturbances.
Warning
:
Using cell phone or microwave oven, HF surgical equipment, magnetic resonance imaging
or other radio radiant equipment near this product may cause malfunction or lead to loss of
essential performance, which means that the measurement accuracy will be affected.
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and
the other equipment should be observed to verify that they are operating normally.
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the
thermometer. Otherwise, degradation of the performance of this equipment could result.
Use of accessories, transducers other than those specified or provided by the manufacturer
of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
Caution:
Security, antitheft, and radiofrequency identification (RFID) devices. Some electromagnetic
anti-theft systems and metal detectors such as those used at entrances or exits of
department stores, libraries, and other public places, and airport security screening devices
may affect the thermometer. Additionally, RFID devices, which are often used to read
identification badges, as well as some tag deactivation devices, such as those used at
payment counters at stores and loan desks at libraries, may also affect the thermometer.
Please do not use thermometer near these places. If you have to go through one of these
devices, turn off your thermometer. Before each usage, checking the status of your
thermometer to ensure it can operating normally.
Using short-wave diathermy, microwave diathermy, or therapeutic ultrasound diathermy (all
now referred to as diathermy) and electrocautery devices near this product may cause
malfunction or lead to loss of essential performance, please do not use thermometer near
these equipment. Before each usage, observing the device to verify that they are operating
normally.
Guidance and manufacture's declaration – electromagnetic emissions
The Infrared thermometer is suitable for use in the specified electromagnetic environment (s)
and it has meets the following standard’s emission requirements.
Phenomenon
Profession healthcare facility
environment
Home healthcare environment
Radiated RF emissions
CISPR 11, Group 1, Class B
CISPR 11, Group 1, Class B
Guidance and manufacturer’s declaration — electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.