  (1)   Set the mode-dial “OFF”.
  (2)   Set 0 kgf

cm indication by means of ZERO-adjustment under no load condition.

  (3)  Charge some torque on the Pick-up and adjust it turning the GAIN-adjust by a 

driver so that the digital indication may accord to the torque.

It holds the peak of input signal and indicate the value. At the point of 
timer-finish, the time-out indication lamp(4) will light. It holds the indicated 
value within the setting time (1~3 sec.), and after that it cancels automati-
cally showing 0 kgf

cm. Usually this mode is applied to measure the 

torque of Impulse Wrenches, Angle Nut Runners, Impact Wrenches, etc.

It indicates the input signals each time at data-indication. Apply this mode 
for ZERO-adjustment and GAIN-adjustment.

8.  GAIN-adjustment  ( How to calibrate )

9.

 

Printer output

        Frame format of one text.

Torque data           Count data        Print  order (L/F)

HOLD1.2.3

1.  Fix the pick-up firmly using M10 bolts on flat or vertical plate or hold it on a vice 

firmly. (For YET-5001C 

10001C, M14 bolts are used.)

2.  In order to connect the amp. to the pick-up, install the connect cable into the connec-

tor(11) and the connector(16).

3.  Connect the input code to the power source and set the power switch(2) on. Lamp 

of power switch will light.

4.  Turn off the Mode dial(3), and turn the ZERO adjust(6) right and left to adjust.

5.  As you can set the measuring time with the unit of 0.1 second upto 9.9 seconds 

under P-P mode by means of the digital switch(5) for setting  time-out, it is possible 
to measure the starting torque.

6.  Time-out indication lamp(4)
     Reaching the setting time set by the time digital switch(5), the time-out lamp(4) will 

light, and stop the rotation of the measuring tool.

7.  Mode dial(3)

HOW TO OPERATE

(For each name, refer to the page 5, exterior drawing.)

P-P mode :

OFF mode :

Содержание YET-501C

Страница 1: ...YET 501C PULSE COUNT POWER MODE TIME OUT TIME ZERO ADJ N m 0 1 1 2 3 OFF P P MODELS YET 501C YET 1001C YET 2001C YET 5001C YET 10001C...

Страница 2: ...1 YET 2001C PULSE COUNT POWER MODE TIME OUT TIME ZERO ADJ N m 0 1 1 2 3 OFF P P 1 2 0 1 9 9 0 1 3 4 R O 0 5 5 6 7 MODELS YET 501C 1001C 2001C 5001C 10001C 0 1 9 9...

Страница 3: ...8sq 1 2sq 20 100 N m 0 1 N m 10 100 0 5 N m 50 200 N m 0 1 N m 20 200 1 N m 100 500 N m 0 2 N m 50 500 2 5 N m 9 9 Kg 10 1 Kg 10 0 Kg 22 7 Kg AC100V 240V 10 50 60 15W 3 0 Kg 1 YET 501C YET 1001C YET...

Страница 4: ...3 1 OFF 2 N m 3 P P 1 3 0 OFF 8 9 1 L F HOLD1 2 3 1 M10 YET 5001C 10001C M14 2 3 ON 4 OFF 0 5 P P MODE 0 1 9 9 6 7 P 5...

Страница 5: ...4 10 RS 232C 4800 bps 1200 2400 9600 bps 1 1bit 8bit 7bit NONE 2bit 1 ASCII No 1 N C 2 TXD 4 RTS 5 CTS 20 DTR 7 S G D Sub 25P C R L F 1 2 3 4 ON 13 1 25 14...

Страница 6: ...100 240V SO 60Hz15W RS 232C SENSOR GAIN ANALOG OUTPUT 250 230 130 YET 5001C 10001C POWER MODE TIME OUT ZERO PULSE COUNT FUSE SENSOR GAIN RS232C 1 8 7 14 2 3 6 5 4 10 15 11 12 13 16 YET 501C 9 52mm 3 8...

Страница 7: ...PULSE COUNT POWER MODE TIME OUT TIME ZERO ADJ N m 0 1 1 2 3 OFF P P Digital Torque Testers MODELS YET 501C YET 1001C YET 2001C YET 5001C YET 10001C I n s t r u c t i o n M a n u a l YOKOTA INDUSTRIAL...

Страница 8: ...r keeps off oil and dust from the indication panel of the amplifier 6 Easy operation using micro computer system 7 Pulse Counter is built in Install the testers in the tool control room the test room...

Страница 9: ...impulse tools and nutrunner only Impact type tools should be measured within 70 of above specified measur ing range 350Nm or less Above measuring range is limited impulse tools and nutrunner only Impa...

Страница 10: ...t Frame format of one text Torque data Count data Print order L F HOLD1 2 3 1 Fix the pick up firmly using M10 bolts on flat or vertical plate or hold it on a vice firmly For YET 5001C 10001C M14 bolt...

Страница 11: ...osition Pin No 1 Disconnection N C 2 TXD Transmission Data 4 RTS 5 CTS 20 DTR Ready mark is always output 7 S G The place for signal The connector on the amplifier side D Sub 25P Use the cross cable w...

Страница 12: ...ont panel 1 Digital display 2 Power switch POWER 3 Mode select switch MODE 4 Time out display lamp TIME OUT 5 Time out setting switch 6 Zero adjust ZERO 7 Dustproof protective cover 8 Pulse count disp...

Отзывы: