6.3.2
Adjustment for STED-7 Type (Universal Type)
The adjustment for STED-7 type without Zero/Span adjustment push-switches on the front
panel is to be performed using JHT200 Handy Terminal.
(a) Connect the test equipment corresponding to each input referring to Figure 6-1 through
Figure 6-3.
(b) Set the parameter write protect (W.P.) of setting jumper to OFF. (refer to “5.2 Setting
Jumper”.)
(c) Turn on the power while the equipment is connected to the instrument, and allow a
warm-up period of about 5 minutes.
(d) Connect JHT200 Handy Terminal.
<Connection>
Adapter for Modular
jack(E9786WH)
BRAIN
connector
JHT200
Handy Terminal
Cable of 5-pin connector
type for communication
(F9182EE)
F0605.ai
Figure
6-5
Connection
• For details of operation and adjusting procedures of JHT200 Handy Terminal, refer
to the instruction manual ‘‘JHT200 Handy Terminal’’ (IM 77J50H01-01EN).
• Do not turn off the power of the instrument during adjustment.
(e) Call the adjustment item (P:ADJUST).
(f) P03:ZERO ADJ1 is displayed.
(g) Apply an input equivalent to 0 % of the input range. Check the input value and the input
display of P03:ZERO ADJ1. If the input value does not correspond to the display value,
select P03:ZERO ADJ1 to enter the adjustment mode.
Mainly select INC (addition) or DEC (subtraction) for adjustment. (Selecting RST resets
the adjusted value and retrieves the factory-set default.) Selecting HINC or HDEC
performs adjustment using a value ten times as large as INC or DEC.
(h) Apply an input equivalent to 100 % of the input range. Check the input value and the
input display of P04:SPAN ADJ1. If the input value does not correspond to the display
value, select P04:SPAN ADJ1 to enter the adjustment mode.
Mainly select INC (addition) or DEC (subtraction) for adjustment. (Selecting RST resets
the adjusted value and retrieves the factory-set default.) Selecting HINC or HDEC
performs adjustment using a value ten times as large as INC or DEC.
(i) After completing the adjustment, set the parameter write protect (W.P.) of setting jumper
to ON. (refer to “5.2 Setting Jumper”.)