12-10
IM WT3001E-51EN
• Setting the Predefined Rated Voltage
(valid when the assignment method is Set)
12.
Press the
cursor keys
to select Un Set.
13.
Press
SET
to display the entry box.
14.
Press the
cursor keys
to set the predefined value.
15.
Press
SET
or
ESC
to close the box.
When the measurement
method is Flicker
When the measurement
method is dmax
• Setting the Measurement Target Frequency
16.
Press the
cursor keys
to select Frequency.
17.
Press
SET
to select 50Hz or 60Hz.
When the measurement
method is Flicker
When the measurement
method is dmax
If the measurement method is set to Flicker (normal flicker measurement), set the
measurement target voltage, single observation period, measurement count, and steady-
state range according to the procedure below.
• Setting the Measurement Target Voltage (IEC 61000-4-15 Ed2.0 Only)
18.
Press the
cursor keys
to select Voltage.
19.
Press
SET
to select 230V or 120V.
• Setting the Single Observation Period
20.
Press the
cursor keys
to select the min or sec box by Interval.
21.
Press
SET
to display the entry box.
22.
Press the
cursor keys
to set the minutes or seconds.
23.
Press
SET
or
ESC
to close the box.
24.
Repeat steps 20 and 23 to enter the minutes and seconds.
12.3 Setting the Measurement Conditions