8.4 Turning ON/OFF the Timeout Warning Display
of the PLL Lock
Procedure
START STOP
REMOTE
CAL
MOTOR
SET
CURSOR
NULL
MENU
STORE SET
MENU
LOWER ITEM
LOWER FORM
ALL
USER SET
DISPLAY
ITEM & ELEMENT
RESET
SET
PAGE
PAGE
SCALING
HRM SET
MEASURE
SYNC SOURCE
AVG
LINE FILTER
FREQ
FILTER
FILE
MISC
STORE
IMAGE SAVE
HELP
NUMERIC
WAVE
OTHERS
ITEM
FORM
U / I / P
WP/q/
TIME
USER
ELEMENT
UPDATE
RATE
HOLD
SINGLE
INTEG
LOCAL
SHIFT
Cursor keys
To exit the menu during operation, press
ESC
located above the
soft keys.
In the procedural explanation below, the phrase “press the
cursor
keys
” may be used. This phrase refers to the procedures for
selecting items and entering values and character strings. For
details on the procedures, see section 3.14 in the
User’s Manual,
IM WT3001E-01EN
.
1.
Press
HRM SET
to display the Harmonics menu.
2.
Press the
PLL Input Warning Msg
soft key to select ON or OFF.
Explanation
• Turning ON/OFF the Timeout Warning Display of the PLL Lock
You can select whether to display a timeout warning of the PLL lock when the PLL
circuit in this instrument cannot detect the PLL source frequency within the detection
period. This function is valid only in wide bandwidth harmonic measurement mode.
• ON: Displays the alert message.
• OFF: Initializes the memory without displaying the alert message.
• PLL Detection Period
• When the PLL source is not set to Smp Clk: 5 s
• When the PLL source is set to Smp Clk:
40 s