8.5 Wide Bandwidth Harmonic Measurement Specifications
Item
Specifications
Accuracy
• When the line filter (5.5 kHz) is ON
Frequency
Voltage and Current
±(reading error + measurement
range error)
Power
±(reading error + measurement
range error)
0.1 Hz ≤ f < 10 Hz
10 Hz ≤ f < 30 Hz
30 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 440 Hz
440 Hz < f ≤ 1 kHz
1 kHz < f ≤ 2.5 kHz
2.5 kHz < f ≤ 3.5 kHz
0.25% of r 0.3% of range
0.25% of r 0.3% of range
0.3% of r 0.05% of range
0.6% of r 0.05% of range
1% of r 0.05% of range
2.5% of r 0.05% of range
8% of r 0.05% of range
0.5% of r 0.4% of range
0.5% of r 0.4% of range
0.45% of r 0.1% of range
1.2% of r 0.1% of range
2% of r 0.1% of range
5% of r 0.15% of range
16% of r 0.15% of range
If the fundamental frequency is between 1 kHz and 2.6 kHz
Add 0.5% of reading to the voltage and current accuracy for frequencies greater than 1 kHz.
Add 1% of reading to the power accuracy for frequencies greater than 1 kHz.
• When the line filter (50 kHz) is ON
Frequency
Voltage and Current
±(reading error + measurement
range error)
Power
±(reading error + measurement
range error)
0
.1 Hz ≤ f < 10 Hz
10 Hz ≤ f < 30 Hz
30 Hz ≤ f ≤ 440 Hz
440 Hz < f ≤ 1 kHz
1 kHz < f ≤ 5 kHz
5 kHz
< f ≤ 10 kHz
0.25% of r 0.3% of range
0.25% of r 0.3% of range
0.3% of r 0.05% of range
0.7% of r 0.05% of range
0.7% of r 0.05% of range
3.0% of r 0.05% of range
0.45% of r 0.4% of range
0.45% of r 0.4% of range
0.45% of r 0.1% of range
1.4% of r 0.1% of range
1.4% of r 0.15% of range
6% of r 0.15% of range
If the fundamental frequency is between 1 kHz and 2.6 kHz
Add 0.5% of reading to the voltage and current accuracy for frequencies greater than 1 kHz.
Add 1% of reading to the power accuracy for frequencies greater than 1 kHz.
• When the line filter is OFF
Frequency
Voltage and Current
±(reading error + measurement
range error)
Power
±(reading error + measurement
range error)
0.1 Hz ≤ f < 10
Hz
10 Hz ≤ f < 30 Hz
30 Hz ≤ f ≤ 1 kHz
1 kHz < f ≤ 10 kHz
10 kHz < f ≤ 55
kHz
0.15% of r 0.3% of range
0.15% of r 0.3% of range
0.1% of r 0.05% of range
0.3% of r 0.05% of range
1% of r 0.2% of range
0.25% of r 0.4% of range
0.25% of r 0.4% of range
0.2% of r 0.1% of range
0.6% of r 0.15% of range
2% of r 0.4% of range
• If the fundamental frequency is between 400 Hz and 1 kHz
Add 1.5% of reading to the voltage and current accuracy for frequencies greater than 10 kHz.
Add 3% of reading to the power accuracy for frequencies greater than 10 kHz.
• If the fundamental frequency is between 1 kHz and 2.6 kHz
Add 0.5% of reading to the voltage and current accuracy for frequencies greater than 1 kHz
and less than or equal to 10 kHz.
Add 7% of reading to the voltage and current accuracy for frequencies greater than 10 kHz.
Add 1% of reading to the power accuracy for frequencies greater than 1 kHz and less than
equal to 10 kHz.
Add 14% of reading to the power accuracy for frequencies greater than 10 kHz.
However, all the items below apply to all tables.
• When the crest factor is set to 3
• When λ (power factor) = 1
• Power figures that exceed 440 Hz are reference values.
• For external current sensor range, add 0.2 mV to the current accuracy and add (0.2 mV/
external current sensor range rating)×100% of range to the power accuracy.
• For direct current input range on the 30-A input element, add 0.2 mA to the current accuracy
and add (0.2 mA/direct current input range rating)×100% of range to the power accuracy.
• For direct current input range on the 2-A input element, add 2 µA to the current accuracy and
add (2 µA/direct current input range rating)×100% of range to the power accuracy.
• For n
th
order component input, add ({n/(m+1)}/50)% of (the n
th
order reading) to the n+m
th
order and n-m
th
order of the voltage and current, and add ({n/(m+1)}/25)% of (the n
th
order
reading) to the n+m
th
order and n-m
th
order of the power.
• Add (n/500)% of reading to the n
th
component of the voltage and current, and add (n/250)% of
reading to the n
th
component of the power.
• Accuracy when the crest factor is 6: The same as when the range is doubled for crest factor 3.
• The accuracy guaranteed range by frequency and voltage/current is the same as the
guaranteed range of normal measurement.