<9. Parameter Summary>
9-3
IM 01W07D01-01EN
Object
ID
Attribute
ID
Label
Description
Default value Handling
(*
1
)
4: WNO
10
Special Cmd
No use
0
R
20
Sensor Location A universal parameter to store the comment that
describes the location. Up to 32 characters can be
used.
---
W
21
Sensor Group
A universal parameter to store the comment that
describes the grouping. Up to 16 characters can be
used.
---
W
22
Sensor Info
A universal parameter to store the comment that
describes the sensor itself. Up to 32 characters can
be used.
---
W
23
LpAEq
A-weighted equivalent sound pressure level from
continuous measurement ( L
Aeq,T
)
---
R
24
LpAEq Status
The status indication of the L
Aeq,T
.
---
R
25
LpCEq
C-weighted equivalent sound pressure level from
continuous measurement ( L
Ceq,T
)
---
R
26
LpCEq Status
The status indication of the L
Ceq,T
.
---
R
27
LpCPeak
C-weighted peak sound pressure level, maximum
sound pressure level during the last sampling
period ( L
Cpeak
)
---
R
28
LpCPeak Status The status indication of the L
Cpeak
.
---
R
29
Octave Band L
Un-weighted sound pressure levels in 1/3 octave
bands from 25 Hz to 630 kHz (=15 values).
If the 1/3 Octave Band Switch is OFF (=0), the
values become zero.
---
R
30
Octave Band H Un-weighted sound pressure levels in 1/3 octave
bands from 800 Hz to 16 kHz (=14 values).
If the 1/3 Octave Band Switch is OFF (=0), the
values become zero.
---
R
31
Octave Band
Status
The status indication of the 1/3 Octave Band L and
H.
---
R
32
Process Variable
Status
The integrated status indication of LpAEq Status,
LpCEq Status, LpCPeak Status and Octave Band
Status.
---
R
33
Measurement
Period
Noise measurement is periodically executed per
this time[sec].
60
W
34
Measurement
Timestamp
The time stamp of noise measurement in TAI
format.
---
R
35
Measurement
Publish Time
Delay
The Elapsed time from the noise measurement to
publish in second.
---
R
36
Microphone
Temperature
The estimated temperature of microphone. The
value is calculated as follows. The unit is defined by
the Microphone Temperature Unit parameter.
Tm= Ca * Ta +Cb
Where:
Tm: Microphone Temperature
Ta: Adapter Temperature (UAPMO)
Ca: Microphone Temperature Coeff A (WNO)
Cb: Microphone Temperature Coeff B (WNO)
---
R
37
Microphone
Temperature
Unit
The unit of the Microphone Temperature is defined
by this parameter.
1001: °C (default)
°C
R
38
Microphone
Temperature
Coeff A
The coefficient of the microphone temperature
calculation.
Refer to Attribute#36.
This value is determined by hardware design.
Please do not rewrite.
1.0
W
Содержание WN100
Страница 1: ...User s Manual WN100 Wireless Sound Level Meter IM 01W07D01 01EN IM 01W07D01 01EN 1st Edition ...
Страница 2: ......
Страница 15: ... 3 Component Names 3 2 IM 01W07D01 01EN Infrared port F0302 ai Figure 3 2 Infrared port ...
Страница 56: ... 10 Connecting Sensor 10 8 IM 01W07D01 01EN 10 6 Infrared Configuration Infrared port F1006 ai ...
Страница 58: ......
Страница 59: ......