<7. Setting Parameters>
7-9
IM 01W07D01-01EN
Table 7.5
Diagnostic Status
Bit
Diagnostic Status Contents
NAMUR
NE107
Category(*
1
)
Alert
Type
Alert
Priority
bit31(MSB)
F:Failure status
---
---
---
bit30
C:Function check status
---
---
---
bit29
O:Out of specification status
---
---
---
bit28
M:Maintenance required status
---
---
---
bit27
Faults in electronics
F
78
15
bit26
Faults in sensor or actuator element
F
77
15
bit25
Installation, calibration problem
C
76
11
bit24
Out of service
C
75
8
bit23
Outside sensor limits
O
74
11
bit22
Environmental conditions out of device specification
O
73
11
bit21
Fault prediction: Maintenance required
M
72
11
bit20
Power is critical low: maintenance need short-term
M
71
15
bit19
Power is low: maintenance need mid-term
M
70
11
bit18
Software update incomplete
C
69
11
bit17
Simulation is active
C
68
8
bit16
Faults due to process influence
F
67
15
bit15
Faults due to non-compliance with specified operating conditions
F
66
15
bit14
Other faults
F
65
15
bit13-bit9
reserved by WCI
---
---
---
bit8-bit1
vendor specific area
---
---
---
bit0
Detail information available (*
2
)
1: available
0: no available
---
---
---
(*1) NAMUR NE107 “Self-Monitoring and Diagnosis of Field Devices”
(*2) The bit0 is always “1: available” as WN100 regardless of the device status.
Содержание WN100
Страница 1: ...User s Manual WN100 Wireless Sound Level Meter IM 01W07D01 01EN IM 01W07D01 01EN 1st Edition ...
Страница 2: ......
Страница 15: ... 3 Component Names 3 2 IM 01W07D01 01EN Infrared port F0302 ai Figure 3 2 Infrared port ...
Страница 56: ... 10 Connecting Sensor 10 8 IM 01W07D01 01EN 10 6 Infrared Configuration Infrared port F1006 ai ...
Страница 58: ......
Страница 59: ......