<7. Setting Parameters>
7-6
IM 01W07D01-01EN
7.3.3 Wireless Communication
The information related the wireless network can be
set and checked with Configurator of Field Wireless
Management Console (FWMC Configurator).
Refer to the User’s Manual of YFGW410 Field
Wireless Management Station (IM 01W02D01-
01EN) as the operation of FWMC Configurator.
The following items must be set to use the wireless
communication at least.
(1) Network Information
Set the network information properly with FWMC
Configurator.
Refer to the User’s Manual of YFGW410 Field
Wireless Management Station (IM 01W02D01-
01EN) as the network information.
(2) Output Data (Sampling Data)
Select the Data to be output (publish) as Sampling
Data on FWMC Configurator.
WN100 supports the output (publish) data shown
Table 7.2.
“Diagnostic Status (DIAG_STATUS)” of UAPMO
block outputs the diagnostics status of WN100.
“Packed Data (Attribute60)” of WNO block outputs
the packed value which includes L
Aeq,T
, L
Ceq,T
, L
Cpeak
,
and 1/3 Octave Band. This data can be viewed
using WN30.
“PV” of AI1, AI2 and AI3 block output L
Aeq,T
, L
Ceq,T
and L
Cpeak
each. (Refer to Figure 7.2)
NOTE
Displayed Name of Output Data on Configuration
is different with Output Data (Sampling Data).
Packed Data is displayed as Attribute60 on
FWMC Configurator.
Table 7.2
Output Data for WN100
Object
Name
Output Data
(Sampling
Data)
Displayed
Name on
FWMC
Configurator
Size
[bytes]
UAPMO
Diagnostic
Status
DIAG_STATUS
4
WNO
Packed Data
Attribute60
53
AI1
PV
PV
5
AI2
PV
PV
5
AI3
PV
PV
5
PV
AI2
L
Aeq,T
L
Ceq,T
L
Cpeak
1/3 octave band
Packed Data (Attribute60)
Compress
PV
AI1
PV
AI3
WNO
Diagnostic Status
UAPMO
Output
(Publish)
Output
(Publish)
Output
(Publish)
Output
(Publish)
Output
(Publish)
F0702.ai
Figure 7.2
Data Transfer and Output Flow
(3) Update Time (Publication Period)
Set “Publication Period (Sec)” on FWMC
Configurator each for “Sampling Data” within the
range of 10 to 600seconds.
NOTE
Do not set the update time less than 10 sec,
other than zero, otherwise WN100 stops
measuring.
When set to zero WN100 will stop output
(publishing).
When set the value under the measurement
period, the same value will be sampled
constantly while the time of the measurement
period is not elapsed.
Refer to “Setting of Measurement Period” of
subsection 7.3.5 as the measurement period.
NOTE
The update time setting affects the battery life.
Shorter the update time shorter the battery life.
Содержание WN100
Страница 1: ...User s Manual WN100 Wireless Sound Level Meter IM 01W07D01 01EN IM 01W07D01 01EN 1st Edition ...
Страница 2: ......
Страница 15: ... 3 Component Names 3 2 IM 01W07D01 01EN Infrared port F0302 ai Figure 3 2 Infrared port ...
Страница 56: ... 10 Connecting Sensor 10 8 IM 01W07D01 01EN 10 6 Infrared Configuration Infrared port F1006 ai ...
Страница 58: ......
Страница 59: ......