<8. Maintenance Schedule>
8-8
IM 12Y18A04-01EN
1st Edition : Oct. 04, 2019-00
Table 8.1
Error messages
Messages
Cause
Solution
Login as Technician or Operator
User has not logged in, and trying
to access restricted operations.
User has to login either as a
Technician or Operator to access
the restricted operations.
User is logged in, but does not
have adequate privileges to access
the restricted operations.
User is not authorized to access
the restricted operations.
Password mismatch, Enter to
continue
Incorrect password has been
entered.
Press “Enter” to cancel the wrong
entry and enable re-entry of correct
password. If password is lost
consult factory for backdoor access
code.
Measurement is in progress
Trying to perform measurement
when it is already under way.
Wait for measurement to complete.
User programmable measuring
frequency
Editing option of analysis frequency
is selected.
Edit the analysis frequency.
User programmable calibration
frequency
If same type of solution is assigned
to two or more inputs.
Edit the calibration frequency.
Two or more inputs are of same
type
If same type of solution is assigned
to two or more input.
Reassign solution to inputs.
Error in date, press F1 to continue Date entered is invalid.
Press F1 and re-enter valid date.
Error in hours, press F1 to continue Hours entered is invalid.
Press F1 and re-enter valid hours.
More than 6 users are not allowed,
delete a user to add further
Technician trying to add more than
6 users.
More than 6 users are not
permitted at any time. To add a
new user after 6, delete an existing
user.
No records to delete.
User is trying to delete Calibration/
Analysis records, when there are
no records stored.
None.
Are you sure you want to delete all
analysis results?
All analysis results? User is trying
to delete all analysis records.
(At least one analysis record is
available in the database.)
None. Deleted records are not
recoverable.
Are you sure you want to delete all
calibration results?
User is trying to delete all
calibration records. (At least one
calibration record is available in the
database.)
None. Deleted records are not
recoverable.
No records to display
User is trying to display records,
and no records are available in the
selected data range.
None.
No records to print
User is trying to print records, and
no records are available in the
selected data range.
None.
Only one record to display
User is trying to view graph
when only one analysis record is
available in the database.
None.
Problem with memory allocation.
Insufficient memory in the system
or system is busy.
Do not try to repeatedly view the
graph.
Try again after some duration
or after performing one more
measurement cycle.