<6. Menu Overview>
6-4
IM 12Y18A04-01EN
1st Edition : Oct. 04, 2019-00
Factory Settings Menu
Allows technician to restore factory defaults, update factory defaults and edit factory
settings.
Instrument Details Menu
Allows viewing of instrument details:
• Software version
• Serial number
• Model number
Analysis Label Change Menu
Allows technician to edit the analysis label.
Printing Options Menu
Allows technician to print the system configuration, analysis parameters and method
parameters. Also allows technician to enable/disable online printing. See Chapter 7 for the
default analyzer parameters listing printouts.
6.6
Hardware Setup Menu
Calibrate Thermistor
Allows technician to calibrate the thermistor. Only required if the factory-installed thermistor
has been serviced or replaced. To calibrate the thermistor, user is required to enter the
reaction block temperature using an external reference thermometer.
Calibrate Pumps
Allows technician to calibrate the flow rate of the pumps.
Calibrate Optics
Allows technician to calibrate the optics.
Calibrate Analog Outputs
Allows technician to calibrate the two analog outputs.
Edit Calibrated Values
Allows technician to edit the calibrated results for fine-tuning.
6.7
Prime Fluidics Menu and New Reagents
Blank
Auto Prime All
This menu runs an automatic priming sequence for all reagents, standards and sample
based on the priming volume specified by the user.
Prime Input Source (Sample/Reagent/...)
Allows technician to selectively prime from the available list of devices.
Priming Volume
Allows technician to change the volume used for priming.
Press + to Navigate to NEXT Page
The new reagents blank is the ZERO intercept or offset. A user entry here will affect the zero
calibration value of the calibration parameters.
Set NEW reagents BLANK
Allows the operator to set the BLANK value for the NEW reagents.This value is usually
supplied with the new batch of reagents and its value may vary between 0 to -10. Default
value is -1.0.
PLEASE NOTE THE (-) SIGN.