<8. Maintenance Schedule>
8-6
IM 12Y18A04-01EN
1st Edition : Oct. 04, 2019-00
8.5
Installing Tubing Kit
The tubing in the WE440 Phosphate Analyzer consists of two parts, the Main Tubing Assembly
(K9705CK) and the Pump Head Tubing Assembly (K9705CL). The two sections of tubing are
connected via luer fittings. The complete tubing layout is shown in below.
STEP 1:
lnstall the tubing channel using the pictures below. Note the the white labels will need to
be on the left side of the analyzer. The tubing channel will connect into the upper receiving
channel.
STEP 2:
Insert vent tube 1 and 2 through vent holes in the bottom left of the enclosure as shown in
the picture.