<7. Terms Definition>
7-1
IM 12Y18A04-01EN
1st Edition : Oct. 04, 2019-00
7. Terms Definition
Terms
Description
Possible Values, Options
ALARM
Type
Assigning a fault type to the alarm.
Low Limit, High Limit, Fluidics Fault,
General Fault, not Assigned
Value
Boundary value for triggering an alarm,
only for low limit and high limit type.
0 to XXXXX (XXXXX Value depends on
precision selected in Setup → Display
parameters → result Decimal Point)
Hysteresis
Value gap or difference before an event
change.
0 to XXXXX (XXXXX Value depends on
precision selected in Setup → Display
parameters → result Decimal Point
RELAY
Up to FOUR spare relays available.
MODE
Allows fault ON or OFF operation.
Normal or Fail Safe
ANALOG OUTPUT
Type
Normal or reverse acting.
0 to 20, 4–20, 20-0 or 20–4
Low
Low concentration value.
High
High concentration value.
EXTERNAL INPUT
Label
User programmable label.
Event
Dry contact event.
Close/Open or Open/Close
Action
Selected function.
Analyze Once, Run/Stop, Calibrate,
Validate, Not assigned
DISPLAY PARAMETERS
Contrast
To adjust the display contrast.
120 to 200
Back Light On-Time
Screen saver time duration for backlight
of display.
0 to 255 minutes, 0 for Continuously On
Display Help
Help message for the user to navigate
between the menus.
Enabled, Disabled
Display Info
Message showing the status of the
current cycle performing.
Enabled, Disabled
Results Decimal Point
Selectable precision to show the analysis
result.
XXXX.X, XXX.XX, XX.XXX, X.XXX,
XXXXX, Auto
Large Fonts Screen
To display the result with large fonts.
Enabled, Disabled
CALIBRATION PARAMETERS
Standard
The concentration of the standard used
for calibration.
0 to XXXXX (XXXXX Value depends on
precision selected in Setup → Display
parameters → result Decimal Point
Mode
Mode represents the calibration type.
Single-Point, Base-Line
Frequency
The time duration between two
calibration cycles.
Every 6Hr, 12Hr, 24Hr, Sun, Mon, Tue,
wed, Thu, Fri, Sat, Manual and Editable
0Hr to 99.99Hr
Start Time
Calibrations start time, except for manual
and editable frequency.
Depends on Time format in Setup →
Time Menu
Slope
The CURRENT Slope value represents
the coefficient that is multiplied by
the absorbance to calculate the
concentration of the /- Intercept
value.
Intercept
Reference and current numbers based
on the base line calibration cycle.
Default value of -1
Calibration Tolerance
Repeat tolerance tested during auto
calibration cycles to accept or reject
automatic calibrations.
Default value of +/- 5%
VALIDATION PARAMETERS
Validation Source
Fluid source for validation.
Sample, Standard, Grab Sample, Not
Assigned
Number of Repeats
Number of times that the validation is to
be performed.
0 to 8