6 - 30
IM 11G02M01-01E
It enables you to set a password to be used when
switching the parameter setting mode to the
maintenance mode.
Description of Password Setting screen
Note) The password set herein must be
managed for safety. Failure to enter
the correct password will not open the
Maintenance mode.
• O
2
correction reference value setting screen
• Password Setting screen
Description of O
2
corection
reference value setting screen
O
2
correction value is calculated from the
following equation by values set herein:
O
2
correction concentration value
21 set value
21 - O
2
concentration
measured value
component
concentration
measured value
=
The value of the fractional part in this calculation
is limited to 4.
If it is 4 or more, it is fixed at 4.
x
?
O
2
correction reference value
: 00 to 19% in 1% step
(initial value: 12%)
Setting range
• Station No. setting screen (option)
Setting range
• Station No. : 00 to 31 (initial No.: 00)
* Please refer to another manual (IM 11G02P01-01E)