6 - 11
IM 11G02M01-01E
5. In the “Auto Calibration component
Selection” screen that appears, select
“enable” or “disable” by pressing the
or
key. After setting, press the
ENT
key.
To close “Setting of Auto Calibration
Component” or to cancel this mode midway,
press the
ESC
key.
A previous screen will return.
To close Auto Calibration
Component setting
ESC
(
)
ENT
End of Auto Calibtation Component Setting
The components which were set to enable is calibrated to zero at once at the time of
auto calibration regardless of setting in 6.2.3 Setting of manual zero calibration .
Caution
Auto calibration is made in the following rules according to the setting.
1. Zero calibration at once of CHs (components) which were set to enable .
2. The span of CHs (components) which were set to enable is calibrated in the ascending
order of CH number.
Example 1. In case all of CH1: CO
2
, CH2: CO, CH3: O
2
, were set to enable .
Zero calibration (at once) of CH1 to CH3 span calibration of CH1 (CO
2
)
span calibration of CH2 (CO) span calibration of CH3 (O
2
)
Example 2. In case, out of CH1: CO
2
, CH2: CO and CH3: O
2
, CH1
(
CO
2
) was set to enable ,
CH2 (CO) was set to enable an CH3 (O
2
) was set to disable .
Zero calibration (at once) of CH1 and 2 span calibration of CH1 (CO
2
)
span calibration of CH2 (CO)
Example