3 - 1
IM 11G02M01-01E
3. INSTALLATION
DANGER
CAUTION
This unit is not explosion-proof type. Do not use it in a place with explosive gases to prevent
explosion, fire or other serious accidents.
• For installation, observe the rule on it given in the instruction manual and select a place
where the weight of gas analyzer can be endured.
Installation at an unsuited place may cause turnover or fall and there is a risk of injury.
• For lifting the gas analyzer, be sure to wear protective gloves. Bare hands may invite an
injury.
• Before transport, fix the casing so that it will not open. Otherwise, the casing may be sepa-
rated and fall to cause an injury.
• The gas analyzer is heavy. It should be transported carefully by two or more persons if
manually required. Otherwise, body may be damaged or injured.
• During installation work, care should be taken to keep the unit free from cable chips or other
foreign objects. Otherwise, it may cause fire, trouble or malfunction of the unit.
3.1 Selection of installation location
To install the analyzer for optimum performance, select a location that meets the following conditions;
(1) Use this instrument indoors.
(2) A vibration-free place
(3) A place which is clean around the analyzer.
(4) Power supply
Voltage rating
;100 V AC to 240 V AC
Allowable range
; 85 to 264 V AC
Frequency
; 50 Hz / 60 Hz
Power consumption; 70 VA max.
Inlet
; Conform to EN60320
Protection Class 1
(5) Operation conditions
Ambient temperature
: -5 to 45
°
C
Ambient humidity
: 90% RH or less, no condensation