<APPENDIX>
App.-2
IM 12E01B30-02EN
1st Edition : Mar. 08, 2023-00
4: After the cleaning time elapses, SV4 opens during the Drainage time to wash the Flow cell,
head tank, and the pipes between head tank and the detector.
5: If “Repeat count of wash” is set to over 2, valve working 2 to 4 is repeated. Right before the
last draining, SV3 closes and the zero water stops.
6: Once the draining for cleaning completes, SV4 closes, SV1 opens, and TB830D starts
feeding a sample.
7: After Flow/Calibration Waiting time” elapses, the detector becomes Recovery time.
8: Afthe the Recovery time elapses, a measurement starts.
n
Automatic calibration (-A3, -A5)
CLOSE
OPEN
CLOSE
OPEN
CLOSE
OPEN
CLOSE
OPEN
Drainage time
Drainage time
Drainage time
Drainage time
Washing time
Flow/Calibration
Waiting time
Calibration
time
Flow/Calibration
Waiting time
Flow/Calibration
Waiting time
Recovery time
Washing
Calibration
Recovery Meas.
Washing time
Flow/Calibration
Waiting time
Meas.
Status
SV1(Sample)
SV2(Wash. water)
SV3(Zero)
SV4(Drain)
1: When sequence starts, Automatic cleaning works first. After the Automatic cleaning, an
automatic calibration starts without a recovery time. Therefore, the steps from 1 to 5 for
“Automatic cleaning (-A3)” are common. When a Drainage time starts after the last Cleaning
time, the screen shows a calibration. This time, SV3 (Zero) stays open.
2: After the Drainage time elapses, SV4 closes and a zero water starts flowing into the
detector.
3: After “Flow/Calibration Waiting time” elapses, a zero calibration starts. The calibration
time in the figure shows the total amount of time of an Auto stabilization check and a zero
calibration which is carried out after the Auto stabilization check (the reading becomes
stable).
4: After the calibration, SV4 opens and the zero water in the Flow cell is drained.
5: After the Drainage time elapses, SV4 closes, SV1 opens, and a sample starts flowing into
the detector.
6: After Flow/Calibration Waiting time elapses, TB830D starts the Recovery time.
7: After the Recovery time elapses, a measurement starts.