<2. INSTALLATION, PIPING, AND WIRING>
2-1
IM 12B01B30-02EN
1st Edition : Mar. 08, 2023-00
2. INSTALLATION, PIPING, AND WIRING
The “TB830D Surface Scattering Turbidity Detector” is shipped after being fully packaged so that
it will not be damaged during transportation. Unpack carefully near the installation location.
The TB830D may also be shipped embedded in a Sampling system.
n
Installation Location
The TB830D should be installed in a location:
• In a building or a cabinet where no direct sunlight or rainwater can get inside.
Direct sunlight may cause an abnormal rise in the temperature inside the instrument and
discoloration or deterioration of resin parts. Also, removing the cover for maintenance during
rain may damage the electrical parts inside the instrument.
Install the instrument in a building or cabinet away from direct sunlight and rainwater.
• With little vibration
Vibration may result in incomplete connections such as external wiring.
• Where corrosive gases are not present.
Corrosive gas can damage electrical component inside the instrument.
• Where humidity of 0 to 90% RH is maintained (No condensation)
No installing in a place with high humidity.
• With little temperature change, close to normal temperature
The ambient temperature must not exceed the range of -5 to 55°C.
If the temperature of the sample is lower than the ambient temperature, it may cause
condensation.
Take protective measures to prevent a sample and the tap water from freezing, if necessary.
• With adequate maintenance space and easy access for maintenance work.
Secure sufficient maintenance space for maintenance such as replacement, cleaning, and
calibration.
• Where the drain is provided
The sample needs to be drained during cleaning and calibration, so install it in a place
where drainage is possible.
• Near the place where the converter is installed (without sampling system)
Consider the length of the connection cable to the converter when installing.