<8. MAINTENANCE>
8-9
IM 12B01B30-02EN
1st Edition : Mar. 08, 2023-00
8.9
Lens cleaning
(1) Stops supplying power to TB830D.
(2) Loosen screw A under the detector and open the top cover (Figure 8.9).
(3) Check whether or not the lenses in the detector are dirty. If they are, wipe them with cotton
swabs. At this time, if the cotton swab is soaked with alcohol, it will become cleaner.
(4) After cleaning the lenses, carry out zero and span calibrations.
(5) End.
Lenses
Screw A
Figure 8.9
Lens positions
8.10 Fuse replacement
If a blown fuse occurs, replace the fuse.
There is one fuse in the converter and one fuse in TB830D. When sampling system -A2, -A3, or
-A5 (with auto cleaning) is selected, there are another two fuses in the relay box.
See (IM 12A01G01-02EN) for how to replace the converter fuse. This section explains the
procedures for replacing fuses of the relay box.
CAUTION
Always disconnect the instrument from Power Supply prior to beginning to replacing the fuse.
CAUTION
Stop supplying power to TB830D before opening the detector top cover. Make sure that the
Power Supply of TB830D is not turned on. Do not touch the terminals while Power Supply of
TB830D is on.