< 7. Calibration >
7-4
IM 12E01B30-02EN
1st Edition : Mar. 08, 2023-00
How to install the check tool
• Loosen the screw A at the bottom of the detector and open the top detector. (Figure 7.2)
• Set the check tool as shown in Figure 7.2.
Figure 7.1
Check tool
Place the check tool in the Flow cell.
Slide it to the right and press it.
Check tools
Screw A
Flow cell
Drain
Figure 7.2
How to install a check tool
CAUTION
Do not lose the check tool supplied with the turbidity detector. The check tool has a turbidity value
unique to the turbidity detector of interest. Therefore, the same check tool cannot be purchased
again. If it is lost, contact Yokogawa. We may provide another check tool prepared in advance.
<How to clean the check tool>
Clean the dust and contaminants on the surface of the check tool.
(1) Cleaning with the attached silicone cloth or washing
If the surface is dusty, use the included silicone cloth to gently wipe it off and remove the
dust. At this time, do not rub hard the surface.
If the dust and contaminants cannot be removed by just wiping it, wash it with tap water.
After cleaning, drain the water well and absorb the remaining water with tissue paper. At this
time, do not rub the surface.