Figure 3-6. System configuration
3-3. Wiring of sensors
3-3-1. General precautions
Generally, transmission of signals from SC sensors is at a low voltage and current level. Thus a lot of care
must be taken to avoid interference. Before connecting sensor cables to the transmitter make sure that
following conditions are met:
– the sensor cables are not mounted in tracks together with high voltage and or power switching cables
– only standard sensor cables or extension cable are used
– the transmitter is mounted within the distance of the sensor cables (max. 10 m) + up to 50m WF10
extension cable.
– the setup is kept flexible for easy insertion and retraction of the sensors in the fitting.
3-3-2. Additional precautions for installations in hazardous areas - Intrinsic safe
Make sure that the total of capacitances and inductances connected to the input terminals of the EXA
SC202S do not exceed the limits given in the certificate.
This sets a limit to the cable and extensions used.
– The intrinsic safe version of the EXA 202 instrument can be mounted in Zone 1.
– The sensors can be installed in Zone 0 or Zone 1 if a safety barrier according to the limits given in the
system certificate is used.
– Ensure that the total of capacitances and inductances connected to the terminals of the EXA SC202 do
not exceed the limits given in the certificate of the safety barrier or distributor.
– The cable used should preferably have a BLUE colour or marking on the outside.
– Installation for (sensors in Zone 0 or 1):
Generally, the distributor with input/output isolation has no external earth connection. If there is an earth
connection on the distributor and the external connection of the transmitter is connected to “protective”
earth, the shield of the 2-wire cable may NOT be connected to “protective” earth at the distributor too.
INPUT
OUTPUT/SUPPLY
Safety Barrier
SC202S only
SENSORS
2,5 or 10 m
DISTRIBUTOR
RECORDER
COMPUTER
HAND HELD
COMMUNICATOR
CURRENT OUTPUT
NO
MODE
YES
ENT
>
FAIL
HOLD
YES NO
ENT
MEASURE
MAN.CAL
DISPLAY
HOLD
YOKOGAWA
MODE
TEMP
AUT.CAL
1 2
1 8 0
100
0
>
Содержание SC202G (S)
Страница 19: ...IM 12D7B3 E E Specifications 2 12...
Страница 71: ...IM 12D7B3 E E 12 1 Test Certificate...