IM 12D7B3-E-E
Parameter setting 5-19
Code Display
Function
Function detail
X
Y
Z
Default values
Communication
60
*COMM.
Communication
Set communication
Off
0
1.0
On
Set communication
On
1
Set communication PH201*B On
2
Communication write enable
0
Write
Communication write protect
1
enable
*ADDR.
Network address
Set address 00 to 15
00
61
*HOUR
Clock setup
Adjust to current date and time using
*MINUT
>, ^ and ENT keys
*SECND
*YEAR
*MONTH
*DAY
62
*ERASE
Erase logbook
Press YES to clear logbook data
63-69
Not used
Code Display
Function
Function detail
X
Y
Z
Default values
General
70
*LOAD
Load defaults
Reset configuration to default values
71-79
Not used
Code Display
Function
Function detail
X
Y
Z
Default values
Test and setup mode
80
*TEST
Test and setup
Built in test functions as detailed in QIS
and Service Manual
mA
mA