HART Communication
46
TI 12A01A01-60EN
ISC450G
Parameter
R/W
Parameter
Value
Unit
Note
Param
[Type: Control]
R/W
NONE
CONDUCT
[Measure: Cond
Only, Cond + Conc]
CONCENT1
[Measure: Conc
Only]
TEMPERATURE
CONCENT
[Measure: Cond +
Conc]
CONDUCT2
[Measure: Cond
Only]
CONCENT2
[Measure: Conc
Only]
Parameters can be
written but cannot be
read correctly. Write
the parameters
at the front panel
display.
PID SP
[Type: Control]
R/W
Numeric value
no limit
PID Rng
[Type: Control]
R/W
Numeric value
greater than
0.000000
PID Dir
[Type: Control]
R/W
direct
reverse
PID MR
[Type: Control (Func: P)]
R/W
Numeric value
0.000000 to
100.000000
%
PID I time
[Type: Control (Func: PI, PID) ]
R/W
Numeric value
1.000000 to
3600.000000
s
PID D time
[Type: Control (Func: PID)]
R/W
Numeric value
0.000000 to
60.000000
s
Analog output
[Type: Control]
R/W
PF
DC
Max freq
[Type: Control (Analog output: PF)]
R/W
Numeric value
1.000000 to
120.000000
p/m
DC time
[Type: Control (Analog output:
DC)]
R/W
Numeric value
1.000000 to
1800.000000
s
Expire time
[Type: Control]
R/W
Numeric value
1.000000 to
1800.000000
s
Func
[Type: Alram]
R
Alarm
Param
[Type: Alarm]
R/W
NONE
CONDUCT
[Measure: Cond
Only, Cond + Conc]
CONCENT1
[Measure: Conc
Only]
TEMPERATURE
CONCENT
[Measure: Cond +
Conc]
CONDUCT2
[Measure: Cond
Only]
CONCENT2
[Measure: Conc
Only]
Parameters can be
written but cannot be
read correctly. Write
the parameters
at the front panel
display.
Alarm Sp
[Type: Alarm]
R/W
Numeric value
no limit
Alarm dir.
[Type: Alarm]
R/W
Low
High
Alarm hyst.
[Type: Alarm]
R/W
Numeric value
greater than
0.000000
Oct. 03, 2017-00