HART Communication
9
TI 12A01A01-60EN
PH450G
Parameter
R/W
Parameter
Value
Unit
Note
Param
[Type: Control]
R/W
NONE
PH [Sensor type:
pH] or [Sensor type:
p H + O R P ( M e a s
type: pH, pH+rH,
pH+ORP)]
TEMPERATURE
ORP [Sensor type:
O R P ] o r [ S e n s o r
t y p e : p H + O R P
(Meas type: ORP,
pH+ORP)]
RH [Sensor type:
pH+ORP (Meas type:
rH, pH+rH)]
PID SP
[Type: Control]
R/W
Numeric value
no limit
pH [Param: PH]
ッ
-C,
ッ
-F [Param:
TEMPERATURE]
mV [Param: ORP]
rH [Param: RH]
Note
PID Rng
[Type: Control]
R/W
Numeric value
greater than
0.001000
PID Dir
[Type: Control]
R/W
direct
reverse
PID MR
[Type: Control (Func: P)]
R/W
Numeric value
0.000000 to
100.000000
%
PID I time
[Type: Control (Func: PI, PID)]
R/W
Numeric value
1.000000 to
3600.000000
s
PID D time
[Type: Control (Func: PID)]
R/W
Numeric value
0.000000 to
60.000000
s
Burn
[Type: Control]
R/W
Off
Low
High
Expire time
[Type: Control]
R/W
Numeric value
0.000000 to
1800.000000
s
Func
[Type: Output]
R/W
Linear
Table
Param
[Type: Output]
R/W
NONE
PH [Sensor type:
pH] or [Sensor type:
p H + O R P ( M e a s
type: pH, pH+rH,
pH+ORP)]
TEMPERATURE
ORP [Sensor type:
O R P ] o r [ S e n s o r
t y p e : p H + O R P
(Meas type: ORP,
pH+ORP)]
RH [Sensor type:
pH+ORP (Meas type:
rH, pH+rH)]
Lin 0%
[Type: Output (Func: Linear)]
R/W
Numeric value
no limit
pH [Param: PH]
ッ
-C,
ッ
-F [Param:
TEMPERATURE]
mV [Param: ORP]
rH [Param: RH]
Note
Lin 100%
[Type: Output (Func: Linear)]
R/W
Numeric value
no limit
Burn
[Type: Output]
R/W
Off
Low
High
Damping time
[Type: Output]
R/W
Numeric value
0.000000 to
3600.000000
s
Func
[Type: Simulate]
R
Simulate
Percentage
[Type: Simulate]
R/W
Numeric value
-10.000000 to
110.000000
%
Oct. 03, 2017-00