HART Communication

8

TI 12A01A01-60EN

PH450G

Parameter

R/W

Parameter 

Value

Unit

Note

Zero/

Slope/

ITP

Zero

[Sensor type: pH] or 

[Sensor type: pH+ORP (Meas 

type: pH, pH+rH, pH+ORP)]

R/W

Numeric value

4.000000 to 

7.222200

pH

[Zero unit: pH]

mV

[Zero unit: mV]

Slope

[Sensor type: pH] or 

[Sensor type: pH+ORP (Meas 

type: pH, pH+rH, pH+ORP)]

R/W

Numeric value

44.439999 to 

62.200001

mV/pH

ITP

[Sensor type: pH] or 

[Sensor type: pH+ORP (Meas 

type: pH, pH+rH, pH+ORP)]

R/W

Numeric value

0.000000 to 

14.000000

pH

ORP zero

[Sensor type: ORP] or 

[Sensor type: pH+ORP (Meas 

type: ORP, pH+ORP)]

R/W

Numeric value

-111.110001 to 

1111.109985

mV

ORP slope

[Sensor type: ORP] or 

[Sensor type: pH+ORP (Meas 

type: ORP, pH+ORP)]

R/W

Numeric value

71.999001 to 

102.220001

%

rH zero

[Sensor type: pH+ORP (Meas 

type: rH, pH+rH)]

R/W

Numeric value

-111.110001 to 

1111.109985

mV

rH slope

[Sensor type: pH+ORP (Meas 

type: rH, pH+rH)]

R/W

Numeric value

71.999001 to 

102.220001

%

Impedance 

setting

Input 1 impedance

R/W

Low

High

Imp. 1 low limit

[Input 1 impedance: Low]

R/W

Numeric value

1000.000000 to 

1000000.000000

kOhm

The use of Built-in 

DTM causes an error 

and disables a value 

to be entered when it 

is outside the range 

1000.0 to 1000.1

Imp. 1 high limit

[Input 1 impedance: Low]

R/W

Numeric value

1000.000000 to 

1000000.000000

kOhm

T h e u s e o f B u i l t -

in DTM causes an 

error and disables a 

value to be entered 

when it is outside 

t h e   r a n g e 1 0 0 0 . 0 

to1000.1.

Input 2 impedance

R/W

Low

High

Imp. 2 low limit

[Input 2 impedance: Low]

R/W

Numeric value

1000.000000 to 

1000000.000000

kOhm

The use of Built-in 

DTM causes an error 

and disables a value 

to be entered when it 

is outside the range 

1000.0 to 1000.1.

Imp. 2 high limit

[Input 2 impedance: Low]

R/W

Numeric value

1000.000000 to 

1000000.000000

kOhm

T h e u s e o f B u i l t -

in DTM causes an 

error and disables a 

value to be entered 

when it is outside 

the range1000.0 to 

1000.1.

Output setup

mA1 setup

Type

R/W

Off

Control

Output

Simulate

Func

[Type: Off]

R

Off

Func

[Type: Control]

R/W

P

PI

PID

Oct. 03, 2017-00

Содержание PH450G

Страница 1: ...3 l Connecting Setup Tool 3 n Functions Available via HART Communication 3 n Operational Precaution 4 l Setting on the PH450G SC450G ISC450G and a setup tool communication tool 4 l Operation with hand...

Страница 2: ...ons specific to the PH450G SC450G ISC450G FieldMate contains the versions of DD files at the time when sold To download the latest version of DD files visit our library by clicking the following link...

Страница 3: ...th a four digit number such as 0102 xxxdd The first two digits 01 in the example above indicate the device revision The following two digits 02 in the example above indicate the DD revision Use the la...

Страница 4: ...handling all the parameters of the PH450G SC450G ISC450G PH450G SC450G ISC450G has a lot of parameters Accessing many parameters may take several minutes n DD Online Parameters for PH450G Device revi...

Страница 5: ...Meas type pH pH rH pH ORP R Sensor mV Sensor type pH pH ORP R mV ORP zero Sensor type ORP or Sensor type pH ORP Meas type ORP pH ORP R mV ORP slope Sensor type ORP or Sensor type pH ORP Meas type ORP...

Страница 6: ...temp R W Numeric value 0 000000 to 100 000000 Temp unit C 3 2 0 0 0 0 0 0 t o 212 000000 Temp unit F C F Note Comp method Sensor type pH or Sensor type pH ORP Meas type pH pH rH pH ORP rH R W None TC...

Страница 7: ...P Zero hi lim Sensor type ORP or Sensor type pH ORP Meas type ORP pH ORP R W Numeric value 0 000000 to 1500 000000 mV ORP Zero lo lim Sensor type ORP or Sensor type pH ORP Meas type ORP pH ORP R W Num...

Страница 8: ...20001 Impedance setting Input 1 impedance R W Low High Imp 1 low limit Input 1 impedance Low R W Numeric value 1000 000000 to 1000000 000000 kOhm The use of Built in DTM causes an error and disables a...

Страница 9: ...00 to 3600 000000 s PID D time Type Control Func PID R W Numeric value 0 000000 to 60 000000 s Burn Type Control R W Off Low High Expire time Type Control R W Numeric value 0 000000 to 1800 000000 s F...

Страница 10: ...l R W direct reverse PID MR Type Control Func P R W Numeric value 0 000000 to 100 000000 PID I time Type Control Func PI PID R W Numeric value 1 000000 to 3600 000000 s PID D time Type Control Func PI...

Страница 11: ...type pH ORP Meas type rH pH rH PID SP Type Control R W Numeric value no limit pH Param PH C F Param TEMPERATURE mV Param ORP rH Param RH Note PID Rng Type Control R W Numeric value g r e a t e r t h...

Страница 12: ...larm R W Numeric value 1 000000 to 1800 000000 s Func Type Fail R W Hard Soft Fail Hard Fail Only The use of Built i n D T M d i s a b l e s Hard Soft Fail to be selected Use a device DTM or set at th...

Страница 13: ...Param TEMPERATURE mV Param ORP rH Param RH Note PID Rng Type Control R W Numeric value greater than 0 001000 PID Dir Type Control R W direct reverse PID MR Type Control Func P R W Numeric value 0 0000...

Страница 14: ...000 s Expire time Type Alarm R W Numeric value 1 000000 to 1800 000000 s Func Type Fail R W Hard Soft Fail Hard Fail Only The use of Built in DTM disables Hard S o f t F a i l t o b e selected Func Ty...

Страница 15: ...Param TEMPERATURE mV Param ORP rH Param RH Note PID Rng Type Control R W Numeric value greater than 0 001000 PID Dir Type Control R W direct reverse PID MR Type Control Func P R W Numeric value 0 0000...

Страница 16: ...R W Numeric value 1 000000 to 1800 000000 s Func Type Fail R W Hard Soft Fail Hard Fail Only The use of Built i n D T M d i s a b l e s Hard Soft Fail to be selected Use a device DTM or set at the fro...

Страница 17: ...m TEMPERATURE mV Param ORP rH Param RH Note PID Rng Type Control R W Numeric value g r e a t e r t h a n 0 001000 PID Dir Type Control R W direct reverse PID MR Type Control Func P R W Numeric value 0...

Страница 18: ...s Expire time Type Alarm R W Numeric value 1 000000 to 1800 000000 s Func Type Fail R W Hard Soft Fail Hard Fail Only The use of Built i n D T M d i s a b l e s disabled Hard Soft Fail to be selected...

Страница 19: ...R W Off Warn Fail Temp too high R W Off Warn Fail Temp too low R W Off Warn Fail Matrix config Sensor type pH or Sensor type pH ORP Meas type pH pH rH pH ORP R W Warn Fail ORP too high Sensor type OR...

Страница 20: ...point PID range Manual reset PID I time PID D time Control type DC period time Max pulse freq Alarm setpoint Alarm direction Alarm hysteresis Alarm delay time Expire time Wash setup Erase logbook R W...

Страница 21: ...e use of Built in DTM may make characters displayed as unintended ones C is replaced with C F is replaced with F Use a device DTM or check at the front panel display n DD Online Parameters for SC450G...

Страница 22: ...m Adj CC R cm m Method SC1 R Method SC2 R Polarization R Ohms R ohm USP R Zoom output mA1 value R mA mA2 value R mA S1 perc R S2 perc R S3 perc R S4 perc R Zoom device Serial number R Software Revisi...

Страница 23: ...R W None TC Nacl Matrix Matrix SC1 Method SC1 Matrix R W Ammonia 0 50ppb Morpholine 0 500 ppb Ammonia 15 30 H2SO4 0 27 H2SO4 39 85 H2SO4 93 100 NaOH 0 15 NaOH 25 50 HCl 0 200 ppb HCl 0 18 Hcl24 44 HNO...

Страница 24: ...250 d Concentration Additional table Measure Conc only R W Disable Enable Conc table unit Measure Conc only R W ppt ppm ppb Output setup mA1 setup Type R W Off Control Output Simulate Func Type Off R...

Страница 25: ...trol R W Numeric value 0 000000 to 1800 000000 s Func Type Output R W Linear Table Param Type Output R W NONE CONDUCT Measure Cond only Cond Conc RESIST1 Measure res only CONCENT1 Measure Conc only TE...

Страница 26: ...SP Type Control R W Numeric value no limit S cm S m Param CONDUCT1 2 p p b P a r a m CONCENT1 O cm O m Param RESIST1 2 C F Param TEMPERATURE P a r a m CONCENT2 Note PID Rng Type Control R W Numeric va...

Страница 27: ...t S cm S m Param CONDUCT1 2 ppb Param CONCENT1 O cm O m Param RESIST1 2 C F Param TEMPERATURE Param CONCENT2 Note Lin 100 Type Output Func Linear R W Numeric value no limit S cm S m Param CONDUCT1 2 p...

Страница 28: ...alue greater than 0 000000 S cm S m Param CONDUCT1 2 p p b P a r a m CONCENT1 O cm O m Param RESIST1 2 C F Param TEMPERATURE P a r a m CONCENT2 Note PID Dir Type Control R W direct reverse PID MR Type...

Страница 29: ...lue greater than 0 000000 S cm S m Param CONDUCT1 2 p p b P a r a m CONCENT1 O cm O m Param RESIST1 2 C F Param TEMPERATURE P a r a m CONCENT2 Note Alarm delay Type Alarm R W Numeric value 1 000000 to...

Страница 30: ...ol R W Numeric value greater than 0 000000 S cm S m Param CONDUCT1 2 p p b P a r a m CONCENT1 O cm O m Param RESIST1 2 C F Param TEMPERATURE P a r a m CONCENT2 PID Dir Type Control R W direct reverse...

Страница 31: ...rm R W Numeric value greater than 0 000000 S cm S m Param CONDUCT1 2 p p b P a r a m CONCENT1 O cm O m Param RESIST1 2 C F Param TEMPERATURE P a r a m CONCENT2 Note Alarm delay Type Alarm R W Numeric...

Страница 32: ...e 1 000000 to 3600 000000 s PID D time Type Control Func PID R W Numeric value 0 000000 to 60 000000 s Analog output Type Control R W PF DC Max freq Type Control Analog output PF R W Numeric value 1 0...

Страница 33: ...am Type Control R W NONE CONDUCT Measure Cond only Cond Conc RESIST1 Measure res only CONCENT1 Measure Conc only TEMPERATURE CONDUCT2 Measure Cond only RESIST2 Measure res only CONCENT2 Measure Conc o...

Страница 34: ...Low High Alarm hyst Type Alarm R W Numeric value greater than 0 000000 Alarm delay Type Alarm R W Numeric value 1 000000 to 1800 000000 Expire time Type Alarm R W Numeric value 1 000000 to 1800 00000...

Страница 35: ...W Numeric value 0 000000 to 2 000000 S Res hi lim Measure Res only R W Numeric value 0 000000 to 10000000 000000 ohm Temp too high R W Off Warn Fail Temp too low R W Off Warn Fail Pol detected R W Off...

Страница 36: ...DTM disables the setting properly Num resp preams R W Numeric value 0 to 255 The use of Built in DTM disables the setting properly Review Model R Distributor R Write protect R Manufacturer R Dev id R...

Страница 37: ...2 value R mA S1 perc R S2 perc R S3 perc R S4 perc R Zoom device Serial number R Software Revision R Device Revision R Logbook Calibration Commissioning Input contact R W Disable Factor 10 mA1 Factor...

Страница 38: ...0 15 NaOH 25 50 User defined 1 User defined 2 TC SC1 Method SC1 TC R W Numeric value 0 000000 to 3 5000000 Temp unit C 0 000000 to 2 000000 Temp unit F C F Note Method SC2 Measure Cond only Conc only...

Страница 39: ...ly Cond Conc CONCENT1 Measure Conc Only TEMPERATURE CONCENT Measure Cond Conc CONDUCT2 Measure Cond Only CONCENT2 Measure Conc Only PID SP Type Control R W Numeric value no limit S cm S m Param CONDUC...

Страница 40: ...ype Output Func Linear R W Numeric value no limit S cm S m Param CONDUCT1 2 p p b P a r a m CONCENT1 C F Param TEMPERATURE P a r a m CONCENT2 Note Lin 100 Type Output Func Linear R W Numeric value no...

Страница 41: ...S m Param CONDUCT1 2 p p b P a r a m CONCENT1 C F Param TEMPERATURE P a r a m CONCENT2 Note PID Dir Type Control R W direct reverse PID MR Type Control Func P R W Numeric value 0 000000 to 100 000000...

Страница 42: ...unc Type Simulate R Simulate Percentage Type Simulate R W Numeric value 0 000000 to 100 000000 S1 setup Type R W Off Control Alarm Fail Simulate HOLD Func Type Off R Off Func Type Control R W P PI PID...

Страница 43: ...Measure Cond Conc CONDUCT2 Measure Cond Only CONCENT2 Measure Conc Only Alarm Sp Type Alarm R W Numeric value no limit Note Alarm dir Type Alarm R W Low High Alarm hyst Type Alarm R W Numeric value g...

Страница 44: ...ter than 0 000000 S cm S m Param CONDUCT1 2 p p b P a r a m CONCENT1 C F Param TEMPERATURE P a r a m CONCENT2 Note PID Dir Type Control R W direct reverse PID MR Type Control Func P R W Numeric value...

Страница 45: ...e greater than 0 000000 S cm S m Param CONDUCT1 2 ppb Param CONCENT1 C F Param TEMPERATURE Param CONCENT2 Note Alarm delay Type Alarm R W Numeric value 1 000000 to 1800 000000 s Expire time Type Alarm...

Страница 46: ...ue 1 000000 to 3600 000000 s PID D time Type Control Func PID R W Numeric value 0 000000 to 60 000000 s Analog output Type Control R W PF DC Max freq Type Control Analog output PF R W Numeric value 1...

Страница 47: ...ff Func Type Cntrol R W P PI PID Param Type Control R W NONE CONDUCT Measure Cond Only Cond Conc CONCENT1 Measure Conc Only TEMPERATURE CONCENT Measure Cond Conc CONDUCT2 Measure Cond Only CONCENT2 Me...

Страница 48: ...hyst Type Alarm R W Numeric value greater than 0 000000 Alarm delay Type Alarm R W Numeric value 1 000000 to 1800 000000 Expire time Type Alarm R W Numeric value 1 000000 to 1800 000000 Func Type Fai...

Страница 49: ...ual reset PID I time PID D time Burn Linear Damping time Expire time Multiple checkbox allows more than one option to be selected Contact logbook R W Contact setup PID setpoint PID range Manual reset...

Страница 50: ...Review Model R Distributor R Write protect R Manufacturer R Dev id R Tag R Descriptor R Message R Date R Universal rev R Fld dev rev R Software rev R Poll addr R Num resp preams R Note Built in DTM ma...

Страница 51: ...HART Communication Manual number TI 12A01A01 60EN Sep 2017 1st Edition Newly published n For further information visit Yokogawa website http www yokogawa com an Yokogawa Electric Corporation 2 9 32 N...

Отзывы: