IM CA71-E
10-7
Method of Calibrator Adjustment
10
Step 4: Apply the CAL FS adjustment setpoint input of each range in
Table 10.2 from the calibration standard to the H and L input
terminals of the calibrator.
Step 5: Pressing the
key confirms the CAL FS adjustment
setpoint. At this point, the
symbol blinks.
Step 6: Pressing the
once again saves the adjustment setpoint in
memory.
NOTE
• Range adjustments are made automatically within the calibrator so that the
LCD reading matches the adjustment setpoint in question given in the table
with reference to the input applied as described above. Therefore, you need
not make any range adjustment with
and
keys.
• Saving the adjustment setpoint results in the overwriting of existing data.
Be extremely careful since the previous adjustment setpoints are cleared.
Step 7: The
symbol stops blinking, causing the calibrator to return
to the state discussed in step 4. Using the function selector
switch, select the next range. By repeating steps 4 to 6, you
can adjust the measurement function assigned to that range.
TIP
Press the
key while holding down the
key. This key operation cancels the
CAL mode. You can use the same key operation to cancel the CAL mode during
calibration, before saving to memory.
10.3 Adjusting Measurement Functions
Содержание HANDY CAL CA51
Страница 1: ...User s Manual Model CA51 71 HANDY CAL Calibrators IM CA71 E 1st Edition Feb 2002 IM CA71 E...
Страница 11: ......
Страница 29: ......
Страница 51: ......
Страница 61: ......
Страница 69: ......
Страница 77: ......
Страница 89: ......
Страница 91: ......
Страница 101: ......
Страница 103: ......